FDAClass II · Medical device · Recalled July 8, 2026
Solar GI System (Water Perfusion) recall — Laborie Medical Technologies Corp
Hazard: Under extended clinical use conditions (e.g., long duration studies), moisture generated within the pneumatic circuit may accumulate inside the system's internal tubing or at the hydrophobic filter and waterproof pressure sensors. Accumulated moisture may affect airflow or pressure measurement performance if not properly removed.
Class II: may cause temporary or medically reversible health consequences.
View the official FDA enforcement record (openFDA) ↗
- Official recall ID
- Z-2669-2026
- Brand
- Laborie Medical Technologies
- Classification
- Class II
- Units affected
- 372 units
- Recall status
- Ongoing
- Initiated
- Voluntary: Firm initiated
Description
Solar GI System (Water Perfusion); Product Name (Catalog Number): Solar GI LGI H2O, 12p, Trolley system (G3-6), Solar GI HRM H2O, 24p, Compact Pole system (G3-7), Solar GI HRM H2O, 24p, Trolley system (G3-8), Solar GI HRM H2O, 36p, Trolley system (G3-12), Solar GI HRAM H2O, 24p, Trolley system (G3-14);
Affected products
Companies & origin
- Recalling firm location
- Portsmouth, NH, United States
- Firm address
- 180 International Dr, N/A, 03801-6837
Identifying details (verbatim from the notice)
- Product codes / lots
- 1. Catalog Number: G3-6; UDI-DI: Primary UDI-DI: 08717775955047; Basic UDI-DI: 87177759500G3G8; All serial numbers; 2. Catalog Number: G3-7; UDI-DI: Primary UDI-DI: 08717775955054; Basic UDI-DI: 87177759500G3G8; All serial numbers; 3. Catalog Number: G3-8; UDI-DI: Primary UDI-DI: 08717775955061; Basic UDI-DI: 87177759500G3G8; All serial numbers; 4. Catalog Number: G3-12; UDI-DI: Primary UDI-DI: 08717775955078; Basic UDI-DI: 87177759500G3G8; All serial numbers; 5. Catalog Number: G3-14; UDI-DI: Primary UDI-DI: 08717775955085; Basic UDI-DI: 87177759500G3G8; All serial numbers;
- Distribution
- Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Bangladesh, Belgium, Canada, China, Costa Rica, Croatia, Cyprus, Czech Republic. Denmark, Egypt, France, Germany, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Jordan, Kuwait, Lebanon, Libya, Lithuania, Malaysia, Namibia, Nepal, Netherlands. Oman, Osterreich, Pakistan, Poland, Portugal, Puerto Rico, Qatar, Russia, Russian Federation, Saudi Arabia, Serbia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Switzerland, Taiwan, Thailand, Tunisia, Turkey, UAE, United Kingdom, and Vietnam.
- Firm notified the public via
- Letter
- FDA event ID
- 98876
- Recall initiated
- May 21, 2026
- FDA classified
- July 2, 2026
- FDA report date
- July 8, 2026