Fresh Recalls

⚠ Data freshness notice: the most recent refresh of NHTSA was incomplete, so a very recent recall may be missing. The stored records may also have changed at the agency since the last successful check; for anything time-critical, confirm with the official sources linked below.

ROTOR ASSY-SW40: recall status

Notice names this product. Beckman Coulter’s ROTOR ASSY-SW40 is named in FDA Class II recall Z-0201-2027. Your particular item is not yet verified. Compare the identifiers and exclusions in the notice before deciding whether it applies. Agency-reported hazard: Optima XE XPN ultracentrifuge rotors may contain a microstructure defect that can cause rotors to break into two or more pieces (bifurcate) during a centrifugation run, resulting in destruction of samples being processed. When used for in vitro diagnostic (IVD) sample preparation, rotor breakage may delay patient test results. Additionally, sample destruction may expose laboratory personnel to biohazardous materials. Identifying details: Serial Numbers: 24U15260 to 26U15764; Optima UDI-DI: A99846 - 15099590416324, A99842 - 15099590415754, A99839 - 15099590415822, A99836 - 15099590415846, A99833 - 15099590415860. Units affected: 92 units. Status: Ongoing.

1. Compare your item

This notice names a product; your particular item is not yet verified. Compare every applicable model, serial, lot, production-date or model-year boundary below.

Product named
ROTOR ASSY-SW40
Company / brand
Beckman Coulter
Lot / serial / date codes
Serial Numbers: 24U15260 to 26U15764; Optima UDI-DI: A99846 - 15099590416324, A99842 - 15099590415754, A99839 - 15099590415822, A99836 - 15099590415846, A99833 - 15099590415860
Distribution
US Nationwide distribution.
Agency product description and boundaries

ROTOR ASSY-SW40, Reference Number 331302, when the rotors are installed in the following centrifuges: Optima XPN-100 (Model Numbers A99846 and A94469), Optima XPN-90 (Model Numbers A99842 and A94468), Optima XPN-80 (Model Numbers A99839 and A95765), Optima XE-100 (Model Numbers A99836 and A94516), and Optima XE-90 (Model Numbers A99833 and A94471)

Still to verify: match the label on your item with the full official notice, including any exclusions, tables and photos. Some identifiers appear only on the agency website and are absent from this data feed. A missing field does not mean all versions are affected.

If your lot or date falls outside a stated range, this notice has not established that your item is affected. This is not a finding about other recalls or product safety.

Brand
Beckman Coulter

Recalls naming this model

FDA Class II · Medical device · October 28, 2026

ROTOR ASSY-SW40 recall — Beckman Coulter, Inc.

Optima XE XPN ultracentrifuge rotors may contain a microstructure defect that can cause rotors to break into two or more pieces (bifurcate) during a centrifugation run, resulting i…

Company / brand: Beckman Coulter recalls

Affected product: ROTOR ASSY-SW40 recall status

Official ID Z-0201-2027 · Official record ↗

See all Beckman Coulter recalls →

Verify with the official sources:

Found something wrong or missing? Report a data error.

Compiled from connected official public-domain government data. Sources last checked November 25, 2026 at 12:00 AM UTC.