BD GasPak EZ CO2 Pouch System: recall status

Notice names this product. Becton Dickinson & Co’s BD GasPak EZ CO2 Pouch System is named in FDA Class II recall Z-1982-2026. Your particular item is not yet verified. Compare the identifiers and exclusions in the notice before deciding whether it applies. Agency-reported hazard: BD identified an issue affecting certain lots of BD GasPak EZ CO2 and BD GasPak EZ Campy Gas Generating Pouch Systems, where gas generating sachets may produce carbon dioxide levels below specification. Identifying details: Catalog Nunber: 260684; UDI-DI: 30382902606841; Lot Numbers: 5303157, 5329857, 5329860.. Units affected: 25,760 units. Status: Ongoing.

1. Compare your item

This notice names a product; your particular item is not yet verified. Compare every applicable model, serial, lot, production-date or model-year boundary below.

Product named
BD GasPak EZ CO2 Pouch System
Company / brand
Becton Dickinson & Co
Lot / serial / date codes
Catalog Nunber: 260684; UDI-DI: 30382902606841; Lot Numbers: 5303157, 5329857, 5329860.
Distribution
Worldwide distribution - US Nationwide and the country of Hong Kong.
Agency product description and boundaries

BD GasPak EZ CO2 Pouch System. Catalog Number: 260684. for microbiologic use

Still to verify: match the label on your item with the full official notice, including any exclusions, tables and photos. Some identifiers appear only on the agency website and are absent from this data feed. A missing field does not mean all versions are affected.

If your lot or date falls outside a stated range, this notice has not established that your item is affected. This is not a finding about other recalls or product safety.

Brand
Becton Dickinson & Co

Recalls naming this model

FDA Class II · Medical device · May 6, 2026

BD GasPak EZ CO2 Pouch System recall — Becton Dickinson & Co.

BD identified an issue affecting certain lots of BD GasPak EZ CO2 and BD GasPak EZ Campy Gas Generating Pouch Systems, where gas generating sachets may produce carbon dioxide level…

Company / brand: Becton Dickinson & Co recalls

Affected product: BD GasPak EZ CO2 Pouch System recall status

Official ID Z-1982-2026 · Official record ↗

See all Becton Dickinson & Co recalls →