Fresh Recalls

INCA (Infant Nasal Cannulae Assy) Neonatal Nasal CPAP: recall status

Notice names this product. CooperSurgical’s INCA (Infant Nasal Cannulae Assy) Neonatal Nasal CPAP is named in FDA Class I recall Z-3140-2026. Your particular item is not yet verified. Compare the identifiers and exclusions in the notice before deciding whether it applies. Agency-reported hazard: Potential to exhibit a loose connection between the flexible tubing and the nasal cannula, during assembly or while in use, which may result in loss of CPAP pressure, possibly reducing the effectiveness of treatment and lead to decreased oxygenation until connection is restored. Identifying details: 1. Model Number: 44-0712. UDI-DI: (01)60888937009516. Lot Code: 550031624. 2. Model Number: 44-0707. UDI-DI: (01)20888937009488. Lot Code: 550021057. 3. Model Number: 44-0709. UDI-DI: (01)20888937009495. Lot Code: 550031623. 4. Model Number: 44-0715. UDI-DI: (01)20888937009525. Lot Code: 550031625.. Units affected: 149 boxes (745 units). Status: Ongoing.

1. Compare your item

This notice names a product; your particular item is not yet verified. Compare every applicable model, serial, lot, production-date or model-year boundary below.

Product named
INCA (Infant Nasal Cannulae Assy) Neonatal Nasal CPAP
Company / brand
CooperSurgical
Lot / serial / date codes
1. Model Number: 44-0712. UDI-DI: (01)60888937009516. Lot Code: 550031624. 2. Model Number: 44-0707. UDI-DI: (01)20888937009488. Lot Code: 550021057. 3. Model Number: 44-0709. UDI-DI: (01)20888937009495. Lot Code: 550031623. 4. Model Number: 44-0715. UDI-DI: (01)20888937009525. Lot Code: 550031625.
Distribution
Nationwide distribution to CT, FL, GA, IA, IL, IN, MD, MI, MS, NJ, NY, OH, PA, VA. International distribution to Brazil and Peru.
Agency product description and boundaries

INCA (Infant Nasal Cannulae Assy) Neonatal Nasal CPAP, Replacement Set. Stay-Flex tubing component. 5 units per box. Model Numbers: 44-0712, 44-0707, 44-0709, 44-0715.

Still to verify: match the label on your item with the full official notice, including any exclusions, tables and photos. Some identifiers appear only on the agency website and are absent from this data feed. A missing field does not mean all versions are affected.

If your lot or date falls outside a stated range, this notice has not established that your item is affected. This is not a finding about other recalls or product safety.

Brand
CooperSurgical

Recalls naming this model

FDA Class I · Medical device · September 30, 2026

INCA (Infant Nasal Cannulae Assy) Neonatal Nasal CPAP recall — CooperSurgical, Inc.

Potential to exhibit a loose connection between the flexible tubing and the nasal cannula, during assembly or while in use, which may result in loss of CPAP pressure, possibly redu…

Company / brand: CooperSurgical recalls

Affected product: INCA (Infant Nasal Cannulae Assy) Neonatal Nasal CPAP recall status

Official ID Z-3140-2026 · Official record ↗

See all CooperSurgical recalls →

Verify with the official sources:

Found something wrong or missing? Report a data error.

Compiled from connected official public-domain government data. Sources last checked October 22, 2026 at 6:00 PM UTC.