Fresh Recalls

Giraffe OmniBed Field Replaceable Unit (FRU): recall status

Notice names this product. GE Medical Systems’s Giraffe OmniBed Field Replaceable Unit (FRU) is named in FDA Class II recall Z-2477-2026. Your particular item is not yet verified. Compare the identifiers and exclusions in the notice before deciding whether it applies. Agency-reported hazard: GE HealthCare has become aware of a potential issue involving the canopy soffit beneath the control panel end of the device. Under certain conditions, for example, impact during transport, one or more of the fasteners that secure the soffit can disengage. If this occurs, a fastener could fall into the patient compartment. In rare circumstances, if multiple fasteners disengage, the canopy soffit can become loose or detach. These conditions could potentially result in patient injury. Identifying details: Serial Numbers: 5452896 801057486 800977619 801093759 5584569 5587625 5367527 5603110 5371240 5497382. Units affected: 265 units. Status: Ongoing.

1. Compare your item

This notice names a product; your particular item is not yet verified. Compare every applicable model, serial, lot, production-date or model-year boundary below.

Product named
Giraffe OmniBed Field Replaceable Unit (FRU)
Company / brand
GE Medical Systems
Lot / serial / date codes
Serial Numbers: 5452896 801057486 800977619 801093759 5584569 5587625 5367527 5603110 5371240 5497382
Distribution
Worldwide - US Nationwide distribution.
Agency product description and boundaries

Giraffe OmniBed Field Replaceable Unit (FRU), Model Numbers 6600-1056-400 and 6600-1461-500

Still to verify: match the label on your item with the full official notice, including any exclusions, tables and photos. Some identifiers appear only on the agency website and are absent from this data feed. A missing field does not mean all versions are affected.

If your lot or date falls outside a stated range, this notice has not established that your item is affected. This is not a finding about other recalls or product safety.

Brand
GE Medical Systems

Recalls naming this model

FDA Class II · Medical device · June 24, 2026

Giraffe OmniBed Field Replaceable Unit (FRU) recall — GE Medical Systems, LLC

GE HealthCare has become aware of a potential issue involving the canopy soffit beneath the control panel end of the device. Under certain conditions, for example, impact during tr…

Company / brand: GE Medical Systems recalls

Affected product: Giraffe OmniBed Field Replaceable Unit (FRU) recall status

Official ID Z-2477-2026 · Official record ↗

See all GE Medical Systems recalls →

Verify with the official sources:

Found something wrong or missing? Report a data error.

Compiled from connected official public-domain government data. Sources last checked October 22, 2026 at 6:00 PM UTC.