Fresh Recalls

OMNIFIT SERIES I ACET: recall status

Notice names this product. Howmedica Osteonics’s OMNIFIT SERIES I ACET is named in FDA Class II recall Z-3196-2026. Your particular item is not yet verified. Compare the identifiers and exclusions in the notice before deciding whether it applies. Agency-reported hazard: Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation. Identifying details: UDI-DI 07613327037883 Lot 13MFT. Units affected: 12,156 US. Status: Ongoing.

1. Compare your item

This notice names a product; your particular item is not yet verified. Compare every applicable model, serial, lot, production-date or model-year boundary below.

Product named
OMNIFIT SERIES I ACET
Company / brand
Howmedica Osteonics
Lot / serial / date codes
UDI-DI 07613327037883 Lot 13MFT
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, LATAM, Europe, Asia Pacific, and South Africa.
Agency product description and boundaries

OMNIFIT SERIES I ACET. INSERT 10¿ 22mm ID Catalog No. 2001-2246 Orthopaedic implant components manufactured from UHMWPE (polyethylene)

Still to verify: match the label on your item with the full official notice, including any exclusions, tables and photos. Some identifiers appear only on the agency website and are absent from this data feed. A missing field does not mean all versions are affected.

If your lot or date falls outside a stated range, this notice has not established that your item is affected. This is not a finding about other recalls or product safety.

Brand
Howmedica Osteonics

Recalls naming this model

FDA Class II · Medical device · September 23, 2026

OMNIFIT SERIES I ACET recall — Howmedica Osteonics Corp.

Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels o…

Company / brand: Howmedica Osteonics recalls

Affected product: OMNIFIT SERIES I ACET recall status

Official ID Z-3196-2026 · Official record ↗

See all Howmedica Osteonics recalls →

Verify with the official sources:

Found something wrong or missing? Report a data error.

Compiled from connected official public-domain government data. Sources last checked October 22, 2026 at 6:00 PM UTC.