Fresh Recalls

SCORPIO C-DOME PATELLA 73-2710 73-2910 SCORPIO U-DOME PATELLA 73-3110 73-3508 73-3708 73-3: recall status

Notice names this product. Howmedica Osteonics’s SCORPIO C-DOME PATELLA 73-2710 73-2910 SCORPIO U-DOME PATELLA 73-3110 73-3508 73-3708 73-3 is named in FDA Class II recall Z-3206-2026. Your particular item is not yet verified. Compare the identifiers and exclusions in the notice before deciding whether it applies. Agency-reported hazard: Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation. Identifying details: 73-2710 UDI-DI 07613327033847 Lot NPPV 73-2910 UDI-DI 07613327033854 Lot 14AGFD 73-3110 UDI-DI 07613327033861 Lots 1L45 4JM5 7EL5 AME3 MK74 NTV7 PYHM RE7J VEWJ 73-3508 UDI-DI 07613327033885 Lots HMM5 NLRE 73-3708 UDI-DI 07613327033892 Lots 3467 3D8D PMRHX1 T6YJ T9ME VWDA YAM7 73-3710 UDI-DI 07613327033908 Lots 3W25 DLYP 73-3910 UDI-DI 07613327033915 Lots 14NEYA 1XED 9Y3T D1M9 D2YP EX8T KRT3 KRT3X1 T325 V8X8 X6YE. Units affected: 12,156 US. Status: Ongoing.

1. Compare your item

This notice names a product; your particular item is not yet verified. Compare every applicable model, serial, lot, production-date or model-year boundary below.

Product named
SCORPIO C-DOME PATELLA 73-2710 73-2910 SCORPIO U-DOME PATELLA 73-3110 73-3508 73-3708 73-3
Company / brand
Howmedica Osteonics
Lot / serial / date codes
73-2710 UDI-DI 07613327033847 Lot NPPV 73-2910 UDI-DI 07613327033854 Lot 14AGFD 73-3110 UDI-DI 07613327033861 Lots 1L45 4JM5 7EL5 AME3 MK74 NTV7 PYHM RE7J VEWJ 73-3508 UDI-DI 07613327033885 Lots HMM5 NLRE 73-3708 UDI-DI 07613327033892 Lots 3467 3D8D PMRHX1 T6YJ T9ME VWDA YAM7 73-3710 UDI-DI 07613327033908 Lots 3W25 DLYP 73-3910 UDI-DI 07613327033915 Lots 14NEYA 1XED 9Y3T D1M9 D2YP EX8T KRT3 KRT3X1 T325 V8X8 X6YE
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, LATAM, Europe, Asia Pacific, and South Africa.
Agency product description and boundaries

SCORPIO C-DOME PATELLA 73-2710 73-2910 SCORPIO U-DOME PATELLA 73-3110 73-3508 73-3708 73-3710 73-3910 Orthopaedic implant components manufactured from UHMWPE (polyethylene)

Still to verify: match the label on your item with the full official notice, including any exclusions, tables and photos. Some identifiers appear only on the agency website and are absent from this data feed. A missing field does not mean all versions are affected.

If your lot or date falls outside a stated range, this notice has not established that your item is affected. This is not a finding about other recalls or product safety.

Brand
Howmedica Osteonics

Recalls naming this model

FDA Class II · Medical device · September 23, 2026

SCORPIO C-DOME PATELLA 73-2710 73-2910 SCORPIO U-DOME PATELLA 73-3110 73-3508 73-3708 73-3 recall — Howmedica Osteonics Corp.

Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels o…

Company / brand: Howmedica Osteonics recalls

Affected product: SCORPIO C-DOME PATELLA 73-2710 73-2910 SCORPIO U-DOME PATELLA 73-3110 73-3508 73-3708 73-3 recall status

Official ID Z-3206-2026 · Official record ↗

See all Howmedica Osteonics recalls →

Verify with the official sources:

Found something wrong or missing? Report a data error.