Fresh Recalls

TRIATHLON HINGE INSERT SIZE 4 11MM 5612-P-411 TRIATHLON HINGE INSERT SIZE 4 13MM 5612-P-41: recall status

Notice names this product. Howmedica Osteonics’s TRIATHLON HINGE INSERT SIZE 4 11MM 5612-P-411 TRIATHLON HINGE INSERT SIZE 4 13MM 5612-P-41 is named in FDA Class II recall Z-3198-2026. Your particular item is not yet verified. Compare the identifiers and exclusions in the notice before deciding whether it applies. Agency-reported hazard: Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation. Identifying details: 5612-P-411 UDI-DI 07613327601893 Lot 360958 5612-P-413 UDI-DI 07613327602012 Lot 354573. Units affected: 12,156 US. Status: Ongoing.

1. Compare your item

This notice names a product; your particular item is not yet verified. Compare every applicable model, serial, lot, production-date or model-year boundary below.

Product named
TRIATHLON HINGE INSERT SIZE 4 11MM 5612-P-411 TRIATHLON HINGE INSERT SIZE 4 13MM 5612-P-41
Company / brand
Howmedica Osteonics
Lot / serial / date codes
5612-P-411 UDI-DI 07613327601893 Lot 360958 5612-P-413 UDI-DI 07613327602012 Lot 354573
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, LATAM, Europe, Asia Pacific, and South Africa.
Agency product description and boundaries

TRIATHLON HINGE INSERT SIZE 4 11MM 5612-P-411 TRIATHLON HINGE INSERT SIZE 4 13MM 5612-P-413 Orthopaedic implant components manufactured from UHMWPE (polyethylene)

Still to verify: match the label on your item with the full official notice, including any exclusions, tables and photos. Some identifiers appear only on the agency website and are absent from this data feed. A missing field does not mean all versions are affected.

If your lot or date falls outside a stated range, this notice has not established that your item is affected. This is not a finding about other recalls or product safety.

Brand
Howmedica Osteonics

Recalls naming this model

FDA Class II · Medical device · September 23, 2026

TRIATHLON HINGE INSERT SIZE 4 11MM 5612-P-411 TRIATHLON HINGE INSERT SIZE 4 13MM 5612-P-41 recall — Howmedica Osteonics Corp.

Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels o…

Company / brand: Howmedica Osteonics recalls

Affected product: TRIATHLON HINGE INSERT SIZE 4 11MM 5612-P-411 TRIATHLON HINGE INSERT SIZE 4 13MM 5612-P-41 recall status

Official ID Z-3198-2026 · Official record ↗

See all Howmedica Osteonics recalls →

Verify with the official sources:

Found something wrong or missing? Report a data error.

Compiled from connected official public-domain government data. Sources last checked October 22, 2026 at 6:00 PM UTC.