Fresh Recalls

ORT300: recall status

Notice names this product. IMRIS Imaging’s ORT300 is named in FDA Class II recall Z-3034-2026. Your particular item is not yet verified. Compare the identifiers and exclusions in the notice before deciding whether it applies. Agency-reported hazard: The linear actuator and rotational lock mechanism components may experience a mechanical failure that prevents the system from transitioning between locked and unlocked state. This may pose a risk of injury to the patient if unintended rotational movement occurs during a procedure. Identifying details: 114093-000 - UDI-DI: 00857534006356, Serial Numbers: 10005695, 10002998, 10002997, 10005456, 10005612, 10004372, 10005613, 10002841, 10011119; 114093-600 - UDI-DI: 00857534006462, Serial Numbers: 10002996, 10004515, 10007411, 10007376. Units affected: 13 units. Status: Ongoing.

1. Compare your item

This notice names a product; your particular item is not yet verified. Compare every applicable model, serial, lot, production-date or model-year boundary below.

Product named
ORT300
Company / brand
IMRIS Imaging
Lot / serial / date codes
114093-000 - UDI-DI: 00857534006356, Serial Numbers: 10005695, 10002998, 10002997, 10005456, 10005612, 10004372, 10005613, 10002841, 10011119; 114093-600 - UDI-DI: 00857534006462, Serial Numbers: 10002996, 10004515, 10007411, 10007376
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, China, Japan, Sweden.
Agency product description and boundaries

ORT300, operating room table, ORT300 Modular Table Main Unit Assy Part Numbers 114093-000 and 114093-600

Still to verify: match the label on your item with the full official notice, including any exclusions, tables and photos. Some identifiers appear only on the agency website and are absent from this data feed. A missing field does not mean all versions are affected.

If your lot or date falls outside a stated range, this notice has not established that your item is affected. This is not a finding about other recalls or product safety.

Brand
IMRIS Imaging

Recalls naming this model

FDA Class II · Medical device · September 2, 2026

ORT300 recall — IMRIS Imaging Inc

The linear actuator and rotational lock mechanism components may experience a mechanical failure that prevents the system from transitioning between locked and unlocked state. This…

Company / brand: IMRIS Imaging recalls

Affected product: ORT300 recall status

Official ID Z-3034-2026 · Official record ↗

See all IMRIS Imaging recalls →

Verify with the official sources:

Found something wrong or missing? Report a data error.

Compiled from connected official public-domain government data. Sources last checked October 22, 2026 at 6:00 PM UTC.