Fresh Recalls

Anti-Torque: recall status

Notice names this product. Innovasis’s Anti-Torque is named in FDA Class II recall Z-2859-2026. Your particular item is not yet verified. Compare the identifiers and exclusions in the notice before deciding whether it applies. Agency-reported hazard: Pedicle screw system intended to provide immobilization/stabilization of spinal segments as adjunct to thoracic/lumbar/sacral spine fusion may have anti-torque instrument with undersized condition which could prevent mating tip from engaging with the pedicle screw tulip and rod during tightening, which may lead to surgical delay, intervention/alternative methods to perform final tightening step. Identifying details: LS-749: UDI-DI: M711BTA2020, M711LS7490, Serial Numbers: 002, 003, 007, 012, 014. Units affected: 5. Status: Ongoing.

1. Compare your item

This notice names a product; your particular item is not yet verified. Compare every applicable model, serial, lot, production-date or model-year boundary below.

Product named
Anti-Torque
Company / brand
Innovasis
Lot / serial / date codes
LS-749: UDI-DI: M711BTA2020, M711LS7490, Serial Numbers: 002, 003, 007, 012, 014
Distribution
US Nationwide distribution in the states of TX, CA.
Agency product description and boundaries

Anti-Torque, Vector, Part: LS-749 part of Vector Pedicle Screw System

Still to verify: match the label on your item with the full official notice, including any exclusions, tables and photos. Some identifiers appear only on the agency website and are absent from this data feed. A missing field does not mean all versions are affected.

If your lot or date falls outside a stated range, this notice has not established that your item is affected. This is not a finding about other recalls or product safety.

Brand
Innovasis

Recalls naming this model

FDA Class II · Medical device · August 12, 2026

Anti-Torque recall — Innovasis, Inc

Pedicle screw system intended to provide immobilization/stabilization of spinal segments as adjunct to thoracic/lumbar/sacral spine fusion may have anti-torque instrument with unde…

Company / brand: Innovasis

Affected product: Anti-Torque

Official ID Z-2859-2026 · Official record ↗

See all Innovasis recalls →