Notice names this product. Insulet’s Omnipod DASH Pod is named in FDA Class I recall Z-2604-2026.
Your particular item is not yet verified. Compare the identifiers and exclusions in the notice before deciding whether it applies. Agency-reported hazard: External soft cannula damage during manufacturing that results in insulin leaking around the Pod instead of being delivered to the user regardless of basal or bolus delivery. The primary failure mode is pump under-delivery due to loss of insulin to an external leak. If insulin is not delivered properly, users may experience high blood glucose levels due to under-delivery of insulin. In the most severe cases, prolonged and persistent high blood glucose levels can lead to diabetic ketoacidosis (DKA), hyperglycemic hyperosmolar syndrome (HHS), or even death. Identifying details: 1. Model/REF Number: POD-BLE-I1-525 (5-Pack), POD-BLE-I1-520 (Single Pod). Pack UDI-DI: 20385082000051. Single Pod UDI-DI: 10385082000023. Lot Numbers: PD1U11052421, PD1U11012421, PD1U11022421, PD1U02152521, PD1U02192521, PD1U04242522, PD1U10282522, PD1U10292521, PD1U10302521, PD1U10312521, PD1U11032521, PD1U11042521, PD1U01082621. 2. Model/REF Number: POD-BLE-I1-529 (10-Pack), POD-BLE-I1-520 (Single Pod). Pack UDI-DI: 20385082000020. Single Pod UDI-DI: 10385082000023. Lot Numbers: PD1U11192421, PD1U09102421, PD1U09092421, PD1U09062421, PD1U10242421, PD1U10252421, PD1U10282421, PD1U10292421, PD1U10312421, PD1U11042421. 3. Model/REF Number: BLE-I1-529 (10-Pack), BLE-I1-520 (Single Pod). Pack UDI-DI: 20385082000020. Single Pod UDI-DI: 10385082000023. Lot Numbers: PD1U11152421, PD1U11182421, PD1U11202421, PD1U09042421, PD1U11132421, PD1U11122421, PD1U11142421, PD1U11152411, PD1U11162421, PD1U04292521, PD1U09032421, PD1U11112421, PD1U04262521. 4. Model/REF Number: BLE-C1-529 (10-Pack), BLE-C1-520 (Single Pod). Pack UDI-DI: 20385082000112. Single Pod UDI-DI: 10385082000115. Lot Numbers: PD1U03042521, PD1U09052421, PD1U11072421, PD1U11082421, PD1U03062521, PD1U03032521, PD1U03072521, PD1U02282525, PD1U02282526, PD1U04252521, PD1U11062521, PD1U11072521, PD1U11102521, PD1U11112511, PD1U11112522, PD1U11122521.. Units affected: 1,790,784 units. Status: Ongoing.
1. Compare your item
This notice names a product; your particular item is not yet verified. Compare every applicable model, serial, lot, production-date or model-year boundary below.
- Product named
- Omnipod DASH Pod
- Company / brand
- Insulet
- Lot / serial / date codes
- 1. Model/REF Number: POD-BLE-I1-525 (5-Pack), POD-BLE-I1-520 (Single Pod). Pack UDI-DI: 20385082000051. Single Pod UDI-DI: 10385082000023. Lot Numbers: PD1U11052421, PD1U11012421, PD1U11022421, PD1U02152521, PD1U02192521, PD1U04242522, PD1U10282522, PD1U10292521, PD1U10302521, PD1U10312521, PD1U11032521, PD1U11042521, PD1U01082621. 2. Model/REF Number: POD-BLE-I1-529 (10-Pack), POD-BLE-I1-520 (Single Pod). Pack UDI-DI: 20385082000020. Single Pod UDI-DI: 10385082000023. Lot Numbers: PD1U11192421, PD1U09102421, PD1U09092421, PD1U09062421, PD1U10242421, PD1U10252421, PD1U10282421, PD1U10292421, PD1U10312421, PD1U11042421. 3. Model/REF Number: BLE-I1-529 (10-Pack), BLE-I1-520 (Single Pod). Pack UDI-DI: 20385082000020. Single Pod UDI-DI: 10385082000023. Lot Numbers: PD1U11152421, PD1U11182421, PD1U11202421, PD1U09042421, PD1U11132421, PD1U11122421, PD1U11142421, PD1U11152411, PD1U11162421, PD1U04292521, PD1U09032421, PD1U11112421, PD1U04262521. 4. Model/REF Number: BLE-C1-529 (10-Pack), BLE-C1-520 (Single Pod). Pack UDI-DI: 20385082000112. Single Pod UDI-DI: 10385082000115. Lot Numbers: PD1U03042521, PD1U09052421, PD1U11072421, PD1U11082421, PD1U03062521, PD1U03032521, PD1U03072521, PD1U02282525, PD1U02282526, PD1U04252521, PD1U11062521, PD1U11072521, PD1U11102521, PD1U11112511, PD1U11112522, PD1U11122521.
- Distribution
- Worldwide distribution - US Nationwide and the countries of Austria, Australia, Belgium, Canada, Switzerland, Germany, Denmark, Finland, France, UK, Greece, Israel, Italy, Kuwait, Leichtenstein, Netherlands, Norway, Saudi Arabia, Sweden and Turkey.
Agency product description and boundaries
Omnipod DASH Pod. Includes the below Model/REF Numbers: 1. POD-BLE-I1-525 (5-Pack), POD-BLE-I1-520 (Single Pod). 2. POD-BLE-I1-529 (10-Pack), POD-BLE-I1-520 (Single Pod). 3. BLE-I1-529 (10-Pack), BLE-I1-520 (Single Pod). 4. BLE-C1-529 (10-Pack), BLE-C1-520 (Single Pod).
Still to verify: match the label on your item with the full official notice, including any exclusions, tables and photos. Some identifiers appear only on the agency website and are absent from this data feed. A missing field does not mean all versions are affected.
If your lot or date falls outside a stated range, this notice has not established that your item is affected. This is not a finding about other recalls or product safety.
FDA Class I · Medical device · July 8, 2026
External soft cannula damage during manufacturing that results in insulin leaking around the Pod instead of being delivered to the user regardless of basal or bolus delivery. The p…
Company / brand: Insulet recalls
Affected product: Omnipod DASH Pod recall status
Official ID Z-2604-2026 ·
Official record ↗
Compiled from connected official public-domain government data. Sources last checked October 22, 2026 at 6:00 PM UTC.