Arvis Hip & Knee 2: recall status

Notice names this product. Kico Knee Innovation Company’s Arvis Hip & Knee 2 is named in FDA Class II recall Z-2293-2026. Your particular item is not yet verified. Compare the identifiers and exclusions in the notice before deciding whether it applies. Agency-reported hazard: The Impactor V-Block Assembly and Trackers A, B, C and E may exhibit signs of early wear and/or corrosion on the exposed magnet face. Identifying details: All Lots/serial Numbers; UDI: IN-28000: 00810180350020; IN-10500: (01)00810180350686; IN-10600 (01)00810180350693; IN-16400:(01)00810180350525; IN-16100: (01)00810180350563;. Units affected: 124 kits. Status: Ongoing.

1. Compare your item

This notice names a product; your particular item is not yet verified. Compare every applicable model, serial, lot, production-date or model-year boundary below.

Product named
Arvis Hip & Knee 2
Company / brand
Kico Knee Innovation Company
Lot / serial / date codes
All Lots/serial Numbers; UDI: IN-28000: 00810180350020; IN-10500: (01)00810180350686; IN-10600 (01)00810180350693; IN-16400:(01)00810180350525; IN-16100: (01)00810180350563;
Distribution
Worldwide - US Nationwide distribution in the states of MS, CA, IA, IN, MI, CO, MT, FL, IL, NY, AL, WI, NE, AZ, MN, VT, NH, MD, RI, MA, NJ, PA, KS, WA, ID, OR, TN, VA, PR and the countries of Australia, France, Germany, Italy.
Agency product description and boundaries

Arvis Hip & Knee 2.0 Instrument Set, REF: IN-21200 containing one of the following components: IN-10500, Impactor V-Block Assembly; IN-10600 , Tracker B; IN-16400, Tracker A; IN-16100, Tracker C

Still to verify: match the label on your item with the full official notice, including any exclusions, tables and photos. Some identifiers appear only on the agency website and are absent from this data feed. A missing field does not mean all versions are affected.

If your lot or date falls outside a stated range, this notice has not established that your item is affected. This is not a finding about other recalls or product safety.

Brand
Kico Knee Innovation Company

Recalls naming this model

FDA Class II · Medical device · June 10, 2026

Arvis Hip & Knee 2 recall — Kico Knee Innovation Company

The Impactor V-Block Assembly and Trackers A, B, C and E may exhibit signs of early wear and/or corrosion on the exposed magnet face.

Company / brand: Kico Knee Innovation Company recalls

Affected product: Arvis Hip & Knee 2 recall status

Official ID Z-2293-2026 · Official record ↗

See all Kico Knee Innovation Company recalls →

Verify with the official sources:

Found something wrong or missing? Report a data error.

Compiled from connected official public-domain government data. Sources last checked October 22, 2026 at 6:00 PM UTC.