Halyard Purezero Ultraviolet Nitrile Exam Gloves Reference Numbers: recall status
Notice names this product. O&M HALYARD’s Halyard Purezero Ultraviolet Nitrile Exam Gloves Reference Numbers is named in FDA Class II recall Z-2230-2026.
Your particular item is not yet verified. Compare the identifiers and exclusions in the notice before deciding whether it applies.Agency-reported hazard: Nitrile Exam Glove failed to meet specifications chemical permeation performance. Identifying details: REF: 48771/UDI: 30680651487712 (Package)20680651487715 (Primary)90680651487714 (Unit of Use) ; 48772/30680651487729 (Package)20680651487722 (Primary)90680651487721 (Unit of Use);48773/30680651487736 (Package)20680651487739 (Primary)90680651487738 (Unit of Use);48774/UDI: 30680651487743 (Package)20680651487746 (Primary) 90680651487745 (Unit of Use);48775/UDI:30680651487750 (Package)20680651487753 (Primary)90680651487752 (Unit of Use). Units affected: 452,500 units. Status: Ongoing.
1. Compare your item
This notice names a product; your particular item is not yet verified. Compare every applicable model, serial, lot, production-date or model-year boundary below.
Still to verify: match the label on your item with the full official notice, including any exclusions, tables and photos. Some identifiers appear only on the agency website and are absent from this data feed. A missing field does not mean all versions are affected.
If your lot or date falls outside a stated range, this notice has not established that your item is affected. This is not a finding about other recalls or product safety.