Fresh Recalls

Brand Name: recall status

Notice names this product. Oxos Medical’s Brand Name is named in FDA Class II recall Z-2857-2026. Your particular item is not yet verified. Compare the identifiers and exclusions in the notice before deciding whether it applies. Agency-reported hazard: Due to a defective circuit board that overheats Identifying details: Lot Code: MC2; UDI 00860008892782; Serial Numbers 1523, 1527, 1488, 1528, 1529, 1544 MC2.1; UDI 00850079606043; Serial Numbers: 1461, 1462, 1547 Full UDI for affected devices: (01)00860008892782(11)260107(21)1523 (01)00860008892782(11)260129(21)1527 (01)00860008892782(11)260109(21)1488 (01)00860008892782(11)260123(21)1528 (01)00860008892782(11)260130(21)1529 (01)00860008892782(11)260227(21)1544 (01)00850079606043(11)260407(21)1461 (01)00850079606043(11)260407(21)1462 (01)00850079606043(11)260408(21)1547. Units affected: 77. Status: Ongoing.

1. Compare your item

This notice names a product; your particular item is not yet verified. Compare every applicable model, serial, lot, production-date or model-year boundary below.

Product named
Brand Name
Company / brand
Oxos Medical
Lot / serial / date codes
Lot Code: MC2; UDI 00860008892782; Serial Numbers 1523, 1527, 1488, 1528, 1529, 1544 MC2.1; UDI 00850079606043; Serial Numbers: 1461, 1462, 1547 Full UDI for affected devices: (01)00860008892782(11)260107(21)1523 (01)00860008892782(11)260129(21)1527 (01)00860008892782(11)260109(21)1488 (01)00860008892782(11)260123(21)1528 (01)00860008892782(11)260130(21)1529 (01)00860008892782(11)260227(21)1544 (01)00850079606043(11)260407(21)1461 (01)00850079606043(11)260407(21)1462 (01)00850079606043(11)260408(21)1547
Distribution
US Nationwide distribution in the states of FL, GA, IN, SC, TX, VA.
Agency product description and boundaries

Brand Name: OXOS Medical, Inc Product Name: MC2 Portable X-Ray System Model/Catalog Number: MC2, MC2.1 Software Version: 5.7.0 Product Description: The MC2 Portable X-Ray System ( MC2 System or MC2 ) is a portable and handheld x-ray system designed to aid clinicians with point of care visualization through diagnostic x-rays of extremities and shoulders. The device allows a clinician to select desired technique factors best suited for their patient anatomy. The MC2 consists of two major system components: the emitter and the cassette. The MC2 emitter and cassette are battery-powered and can be charged with either a wired or wireless charger. The system is intended to interface wirelessly to an external tablet or display when used with the OXOS Device App. The MC2 utilizes an infrared Tracking System and Viewfinder that allows the operator to position the emitter within a safe source-to-image distance (SID) and allows the operator to visualize the anatomy and x-ray field projection prior to initiating an x-ray acquisition. The MC2 also utilizes a LIDAR system to ensure patient safety by maintaining a safe source-to-skin distance. Component: The MC2 Portable X-Ray System is not a component.

Still to verify: match the label on your item with the full official notice, including any exclusions, tables and photos. Some identifiers appear only on the agency website and are absent from this data feed. A missing field does not mean all versions are affected.

If your lot or date falls outside a stated range, this notice has not established that your item is affected. This is not a finding about other recalls or product safety.

Brand
Oxos Medical

Recalls naming this model

FDA Class II · Medical device · August 12, 2026

Brand Name recall — Oxos Medical Inc

Due to a defective circuit board that overheats

Company / brand: Oxos Medical

Affected product: Brand Name

Official ID Z-2857-2026 · Official record ↗

See all Oxos Medical recalls →

Verify with the official sources:

Found something wrong or missing? Report a data error.