Notice names this product. Paragonix Technologies’s Paragonix Technologies Inc is named in FDA Class II recall Z-2967-2026.
Your particular item is not yet verified. Compare the identifiers and exclusions in the notice before deciding whether it applies.Agency-reported hazard: Due to a firmware defect, the portable organ transfer device datalogger may lose its configuration, resulting in a blank screen and loss of displayed preservation parameters. Identifying details: All Serial Numbers/UDI: 00850005470342 Software Version: Paragonix App Ver. 6.1.1/6.1.2. Units affected: 1445 systems. Status: Ongoing.
1. Compare your item
This notice names a product; your particular item is not yet verified. Compare every applicable model, serial, lot, production-date or model-year boundary below.
Product named
Paragonix Technologies Inc
Company / brand
Paragonix Technologies
Lot / serial / date codes
All Serial Numbers/UDI: 00850005470342 Software Version: Paragonix App Ver. 6.1.1/6.1.2
Distribution
US Nationwide distribution in the states and territories of AL, AZ, CA, TN, FL, HI, MN, NJ, TX, UT, CO, CT, DC, MD, GA, KY, LA, MA, MI, MO, MS, NC, ND, NE, NM, NV, NY, PA, PR, SC, UT, VA, WI, WV, IN, OH, KY, IL, MD, MN, OR.
Agency product description and boundaries
Paragonix Technologies Inc. KidneyVault Portable Renal Perfusion System REF# PRGNX-8008-001
Still to verify: match the label on your item with the full official notice, including any exclusions, tables and photos. Some identifiers appear only on the agency website and are absent from this data feed. A missing field does not mean all versions are affected.
If your lot or date falls outside a stated range, this notice has not established that your item is affected. This is not a finding about other recalls or product safety.
Due to a firmware defect, the portable organ transfer device datalogger may lose its configuration, resulting in a blank screen and loss of displayed preservation parameters.