Ingenuity CT CT5000 Ingenuity Plus CT5000 Ingenuity Pro CT5000 Ingenuity Premium Ingenuity: recall status

Notice names this product. PHILIPS MEDICAL SYSTEMS’s Ingenuity CT CT5000 Ingenuity Plus CT5000 Ingenuity Pro CT5000 Ingenuity Premium Ingenuity is named in FDA Class II recall Z-2652-2026. Your particular item is not yet verified. Compare the identifiers and exclusions in the notice before deciding whether it applies. Agency-reported hazard: The clinical CT scan may not automatically trigger after the Bolus Tracking scan and may need to be triggered manually. Additional Bolus Tracking shots may occur. This may result in the clinical acquisition being inadequate for diagnosis, causing a potential for patient rescan and receiving additional radiation. Identifying details: FCO72800815, FCO72800816, 728321 DI (01)00884838059498 728323 DI (01)00884838059504 728326 DI (01)00884838059511 728327 DI (01)00884838098701 728324 DI (01)00884838059863 728325. Units affected: 182. Status: Ongoing.

1. Compare your item

This notice names a product; your particular item is not yet verified. Compare every applicable model, serial, lot, production-date or model-year boundary below.

Product named
Ingenuity CT CT5000 Ingenuity Plus CT5000 Ingenuity Pro CT5000 Ingenuity Premium Ingenuity
Company / brand
PHILIPS MEDICAL SYSTEMS
Lot / serial / date codes
FCO72800815, FCO72800816, 728321 DI (01)00884838059498 728323 DI (01)00884838059504 728326 DI (01)00884838059511 728327 DI (01)00884838098701 728324 DI (01)00884838059863 728325
Distribution
US Nationwide Distribution including Puerto Rico
Agency product description and boundaries

Ingenuity CT CT5000 Ingenuity Plus CT5000 Ingenuity Pro CT5000 Ingenuity Premium Ingenuity CT Upgrades Ingenuity CT China Ingenuity Core 128 Elite Brazil

Still to verify: match the label on your item with the full official notice, including any exclusions, tables and photos. Some identifiers appear only on the agency website and are absent from this data feed. A missing field does not mean all versions are affected.

If your lot or date falls outside a stated range, this notice has not established that your item is affected. This is not a finding about other recalls or product safety.

Brand
PHILIPS MEDICAL SYSTEMS

Recalls naming this model

FDA Class II · Medical device · August 5, 2026

Ingenuity CT CT5000 Ingenuity Plus CT5000 Ingenuity Pro CT5000 Ingenuity Premium Ingenuity recall — PHILIPS MEDICAL SYSTEMS

The clinical CT scan may not automatically trigger after the Bolus Tracking scan and may need to be triggered manually. Additional Bolus Tracking shots may occur. This may result i…

Company / brand: PHILIPS MEDICAL SYSTEMS

Affected product: Ingenuity CT CT5000 Ingenuity Plus CT5000 Ingenuity Pro CT5000 Ingenuity Premium Ingenuity

Official ID Z-2652-2026 · Official record ↗