Notice names this product. PHILIPS MEDICAL SYSTEMS NEDERLAND B.V’s Integris-Allura system is named in FDA Class II recall Z-2229-2026.
Your particular item is not yet verified. Compare the identifiers and exclusions in the notice before deciding whether it applies.Agency-reported hazard: The deaeration hose in X-ray tube cooling units type CU3101, manufactured between February 2016 and May 2020, may degrade over time, potentially resulting in oil leakage. Oil leakage may affect the cooling performance of the X-ray tube. When the oil flow drops below a certain threshold, the system automatically switches to low-dose fluoroscopy and displays the message Low load fluoroscopy flavor selected: Tube cooler problem . This condition cannot be resolved by restarting the system. Identifying details: 1. Integris CV Cesar-Powerpack-Visub-Nicol; Model: 722030; UDI-DI: N/A; Serial Numbers: 4888270, 59, 0, 33; 2. Integris Allura 15 & 12 (monoplane) (722043); UDI-DI: N/A; Serial Numbers: 365, 18151, 176, 239, 407, 426, 0; 3. INTEGRIS Allura 15-12 (biplane) (722044); UDI-DI: N/A; Serial Numbers: 86322, 30, 66; 4. Integris-Allura 9 (722018); UDI-DI: N/A; Serial Numbers: 26, 85; 5. Integris-Allura 9 (Biplane) (722021); UDI-DI: N/A; Serial numbers: 22, 75, 25, 108; 6. Integris H5000 C / Allura 9C (722016); UDI-DI: N/A; Serial Numbers: 4059001008, 4807240, S01H003553/000001;. Units affected: 25 units (11 US, 14 OUS). Status: Ongoing.
1. Compare your item
This notice names a product; your particular item is not yet verified. Compare every applicable model, serial, lot, production-date or model-year boundary below.
Product named
Integris-Allura system
Company / brand
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V
Lot / serial / date codes
1. Integris CV Cesar-Powerpack-Visub-Nicol; Model: 722030; UDI-DI: N/A; Serial Numbers: 4888270, 59, 0, 33; 2. Integris Allura 15 & 12 (monoplane) (722043); UDI-DI: N/A; Serial Numbers: 365, 18151, 176, 239, 407, 426, 0; 3. INTEGRIS Allura 15-12 (biplane) (722044); UDI-DI: N/A; Serial Numbers: 86322, 30, 66; 4. Integris-Allura 9 (722018); UDI-DI: N/A; Serial Numbers: 26, 85; 5. Integris-Allura 9 (Biplane) (722021); UDI-DI: N/A; Serial numbers: 22, 75, 25, 108; 6. Integris H5000 C / Allura 9C (722016); UDI-DI: N/A; Serial Numbers: 4059001008, 4807240, S01H003553/000001;
Distribution
Domestic: Nationwide; International: Canada, Algeria, Argentina, Australia, Austria, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Brunei Darussalam, Bulgaria, Chile, China, Colombia, Costa Rica, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Ethiopia, Finland, France, French Polynesia, Georgia, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Korea, Republic of, Kosovo, Kuwait, Kyrgyzstan, Lao People's Democratic Republic, Latvia, Lebanon, Libya, Lithuania, Malaysia, Maldives, Malta, Martinique, Mexico, Moldova, Mongolia, Morocco, Myanmar, Nepal, Netherlands, New Caledonia, New Zealand, Nigeria, Norway, Oman, Pakistan, Palestine, State of, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, R¿union, Romania, Russian Federation, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Taiwan, Thailand, Trinidad and Tobago, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Viet Nam, Yemen;
Agency product description and boundaries
Integris-Allura system; System Code Description (Model Numbers): Integris CV Cesar-Powerpack-Visub-Nicol (722030), Integris Allura 15 & 12 (monoplane) (722043), INTEGRIS Allura 15-12 (biplane) (722044), Integris-Allura 9 (722018), Integris-Allura 9 (Biplane) (722021), Integris CV Cesar-Powerpack (722030), Integris H5000 C / Allura 9C (722016);
Still to verify: match the label on your item with the full official notice, including any exclusions, tables and photos. Some identifiers appear only on the agency website and are absent from this data feed. A missing field does not mean all versions are affected.
If your lot or date falls outside a stated range, this notice has not established that your item is affected. This is not a finding about other recalls or product safety.
The deaeration hose in X-ray tube cooling units type CU3101, manufactured between February 2016 and May 2020, may degrade over time, potentially resulting in oil leakage. Oil leaka…