Fresh Recalls

Philips MultiDiagnost Eleva with the following Model Numbers: recall status

Notice names this product. PHILIPS MEDICAL SYSTEMS NEDERLAND B.V’s Philips MultiDiagnost Eleva with the following Model Numbers is named in FDA Class II recall Z-2411-2026. Your particular item is not yet verified. Compare the identifiers and exclusions in the notice before deciding whether it applies. Agency-reported hazard: Philips Allura Xper and MultiDiagnost-Eleva systems Hand switch button not fully releasing potentially leading to loss of imaging functionality or unintended radiation exposure and additional contrast injection Identifying details: All codes. Status: Ongoing.

1. Compare your item

This notice names a product; your particular item is not yet verified. Compare every applicable model, serial, lot, production-date or model-year boundary below.

Product named
Philips MultiDiagnost Eleva with the following Model Numbers
Company / brand
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V
Lot / serial / date codes
All codes
Distribution
US and ROW.
Agency product description and boundaries

Philips MultiDiagnost Eleva with the following Model Numbers: Model # 708032 for MultiDiagnost Eleva; Model # 708034 for MultiDiagnost Eleva with Flat Detector; Model # 708036 for MultiDiagnost Eleva; Model # 708037 for MultiDiagnost Eleva with Flat Detector; Model # 708038 for MultiDiagnost Eleva with Flat Detector.

Still to verify: match the label on your item with the full official notice, including any exclusions, tables and photos. Some identifiers appear only on the agency website and are absent from this data feed. A missing field does not mean all versions are affected.

If your lot or date falls outside a stated range, this notice has not established that your item is affected. This is not a finding about other recalls or product safety.

Brand
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V

Recalls naming this model

FDA Class II · Medical device · June 24, 2026

Philips MultiDiagnost Eleva with the following Model Numbers recall — PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.

Philips Allura Xper and MultiDiagnost-Eleva systems Hand switch button not fully releasing potentially leading to loss of imaging functionality or unintended radiation exposure and…

Company / brand: PHILIPS MEDICAL SYSTEMS NEDERLAND B.V

Affected product: Philips MultiDiagnost Eleva with the following Model Numbers

Official ID Z-2411-2026 · Official record ↗

See all PHILIPS MEDICAL SYSTEMS NEDERLAND B.V recalls →

Verify with the official sources:

Found something wrong or missing? Report a data error.