Notice names this product. Philips North America’s MR 5300 is named in FDA Class II recall Z-3312-2026.
Your particular item is not yet verified. Compare the identifiers and exclusions in the notice before deciding whether it applies. Agency-reported hazard: When exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced MR DICOM export is used in conjunction with third-party software for tractography (e.g. for mapping nerve fiber pathways in the brain), the software will not be able to correctly process the DTI data for reconstruction. Identifying details: 1. Model number (REF): 782110; UDI-DI: 00884838099364; Serial Numbers: 63007, 63027, 63064, 63076, 63097, 63123, 63132, 63137, 63148, 63150, 63160, 63236, 63237, 63238, 63258, 63300, 63302, 63360, 63370, 63394, 63397, 63398, 63445, 63523, 63576, 63599, 63636, 63696, 63735, 63766, 63767, 63774, 63806; 2. Model Number (REF): 782152; UDI-DI: 00884838112841; Serial Numbers: 69000, 69007, 69008, 69012, 69015, 69016, 69017, 69030, 69048, 69067, 69068, 69070, 69071, 69076, 69078, 69089, 69094, 69096, 69097, 69101, 69102, 69110, 69121, 69124, 69126, 69128, 69134, 69135, 69137, 69143, 69150, 69154, 69160, 69161, 69171, 69175, 69180, 69193, 69194, 69198, 69209, 69212, 69238, 69239, 69240, 69241, 69247, 69249, 69250, 69263, 69264, 69265, 69266, 69287, 69289, 69290, 69299, 69300, 69304, 69307, 69309, 69310, 69312, 69316, 69318, 69319, 69320, 69321, 69338, 69342, 69344, 69345, 69366, 69367; 3. Model Number (REF): 782156; UDI-DI: 00884838115675; Serial Numbers: 93000, 93007, 93008, 93018, 93023, 93026, 93028, 93037, 93054, 93073, 93084, 93100, 93101, 93102, 93103, 93104, 93105, 93106; 4. Model Number (REF): 782196; UDI-DI: 00884838128446; Serial Numbers: 54016, 54024, 54036, 54040, 54050, 54054;. Units affected: 131. Status: Ongoing.
Compiled from connected official public-domain government data. Sources last checked October 22, 2026 at 12:00 PM UTC.