Notice names this product. Philips North America’s MR 7700 and Upgrade to MR 7700 is named in FDA Class II recall Z-3242-2026.
Your particular item is not yet verified. Compare the identifiers and exclusions in the notice before deciding whether it applies.Agency-reported hazard: A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory. Identifying details: 1. Model Number (REF): 782120; UDI-DI: 884838104112; Serial Numbers: 65002, 65003, 65014, 65015, 65050, 65029, 65056, 65018, 65021, 65035, 65038, 65000, 65025, 65006, 65008, 65043; 2. Model Number (REF): 782153; UDI-DI: 884838112858; Serial Numbers: 68011, 68022, 68031, 68014, 68012, 68009, 68033, 68053, 68023; 3. Model Number (REF): 782130; UDI-DI: 884838104402; Serial Numbers: 45007, 45052, 45043, 45044, 45552, 71616, 42164, 42066, 42104, 42248, 45263, 71532;. Units affected: 37. Status: Ongoing.
1. Compare your item
This notice names a product; your particular item is not yet verified. Compare every applicable model, serial, lot, production-date or model-year boundary below.
Product named
MR 7700 and Upgrade to MR 7700
Company / brand
Philips North America
Lot / serial / date codes
1. Model Number (REF): 782120; UDI-DI: 884838104112; Serial Numbers: 65002, 65003, 65014, 65015, 65050, 65029, 65056, 65018, 65021, 65035, 65038, 65000, 65025, 65006, 65008, 65043; 2. Model Number (REF): 782153; UDI-DI: 884838112858; Serial Numbers: 68011, 68022, 68031, 68014, 68012, 68009, 68033, 68053, 68023; 3. Model Number (REF): 782130; UDI-DI: 884838104402; Serial Numbers: 45007, 45052, 45043, 45044, 45552, 71616, 42164, 42066, 42104, 42248, 45263, 71532;
Distribution
Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Andorra, Angola, Argentina, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belarus, Belgium, Belize, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, C¿te D'Ivoire, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Equatorial Guinea, Estonia, Fiji, Finland, France, French Polynesia, Georgia, Germany, Ghana, Greece, Guadeloupe, Guernsey, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jersey, Jordan, Kazakhstan, Kenya, Korea, Republic of, Kuwait, Kyrgyzstan, Lao People's Democratic Republic, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Macao, Macedonia, Madagascar, Malaysia, Maldives, Malta, Martinique, Mauritius, Mayotte, Mexico, Moldova, Monaco, Mongolia, Montenegro, Morocco, Myanmar, Nepal, Netherlands, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Palestine, State of, Panama, Papua New Guinea, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, R¿union, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tanzania, Thailand, Trinidad and Tobago, Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, Venezuela, Viet Nam, Yemen, Zambia;
Agency product description and boundaries
MR 7700 and Upgrade to MR 7700; Model Number (REF): 1. 782120, 2. 782153, 3. 782130 (Upgrade);
Still to verify: match the label on your item with the full official notice, including any exclusions, tables and photos. Some identifiers appear only on the agency website and are absent from this data feed. A missing field does not mean all versions are affected.
If your lot or date falls outside a stated range, this notice has not established that your item is affected. This is not a finding about other recalls or product safety.
A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the…