Fresh Recalls

Brand Name: recall status

Notice names this product. PTW FREIBURG’s Brand Name is named in FDA Class II recall Z-3174-2026. Your particular item is not yet verified. Compare the identifiers and exclusions in the notice before deciding whether it applies. Agency-reported hazard: When a bolus is included in a treatment plan and sent for secondary dose calculation, the software accounts for it but does not display this clearly, which may lead users to incorrectly omit the bolus during treatment and deliver the prescribed dose to the wrong tissue depth. Identifying details: Model/Catalog Number: S070031.520; UDI-DI: +EPTWS070031Q5200; Software Version 3.1.18;. Units affected: 2. Status: Ongoing.

1. Compare your item

This notice names a product; your particular item is not yet verified. Compare every applicable model, serial, lot, production-date or model-year boundary below.

Product named
Brand Name
Company / brand
PTW FREIBURG
Lot / serial / date codes
Model/Catalog Number: S070031.520; UDI-DI: +EPTWS070031Q5200; Software Version 3.1.18;
Distribution
Worldwide - US Nationwide distribution in the states of GA, NY and the countries of Austria, Belgium, Switzerland, Czechia, Germany, Spain, France, Netherlands, Norway, Poland, Portugal, Romania, T¿rkiye, Chile, United Arab Emirates, Saudi Arabia, Mexico, Hong Kong, Brazil, Tajikistan, Ecuador, Iran, Thailand, Serbia, South Africa, Kazakhstan, Algeria.
Agency product description and boundaries

Brand Name: VERIQA; Product Name: RT MonteCarlo 3D v3; Model/Catalog Number: S070031.520; Software Version: 3.1.18;

Still to verify: match the label on your item with the full official notice, including any exclusions, tables and photos. Some identifiers appear only on the agency website and are absent from this data feed. A missing field does not mean all versions are affected.

If your lot or date falls outside a stated range, this notice has not established that your item is affected. This is not a finding about other recalls or product safety.

Brand
PTW FREIBURG

Recalls naming this model

FDA Class II · Medical device · September 23, 2026

Brand Name recall — PTW FREIBURG

When a bolus is included in a treatment plan and sent for secondary dose calculation, the software accounts for it but does not display this clearly, which may lead users to incorr…

Company / brand: PTW FREIBURG recalls

Affected product: Brand Name recall status

Official ID Z-3174-2026 · Official record ↗

See all PTW FREIBURG recalls →

Verify with the official sources:

Found something wrong or missing? Report a data error.

Compiled from connected official public-domain government data. Sources last checked October 22, 2026 at 6:00 PM UTC.