Notice names this product. Remel’s Yeastone Broth is named in FDA Class II recall Z-0355-2027.
Your particular item is not yet verified. Compare the identifiers and exclusions in the notice before deciding whether it applies.Agency-reported hazard: Affected lot did not meet specifications. Product may report false susceptibility to antifungal agents, specifically within the Azole family. Identifying details: YY3462 lot 440400 UDI-DI 848838018817. Units affected: 33 US, 54 OUS. Status: Ongoing.
1. Compare your item
This notice names a product; your particular item is not yet verified. Compare every applicable model, serial, lot, production-date or model-year boundary below.
Product named
Yeastone Broth
Company / brand
Remel
Lot / serial / date codes
YY3462 lot 440400 UDI-DI 848838018817
Distribution
Worldwide - US Nationwide distribution in the states of MA and CA. The countries of India, Saudi Arabia, UK.
Agency product description and boundaries
Yeastone Broth, 11ML, 10/BOX, used in antibiotic susceptibility testing of mycological clinical specimens
Still to verify: match the label on your item with the full official notice, including any exclusions, tables and photos. Some identifiers appear only on the agency website and are absent from this data feed. A missing field does not mean all versions are affected.
If your lot or date falls outside a stated range, this notice has not established that your item is affected. This is not a finding about other recalls or product safety.