Kinos Axiom Total Ankle System with Axiom PSR: recall status
Notice names this product. restor3d’s Kinos Axiom Total Ankle System with Axiom PSR is named in FDA Class II recall Z-3178-2026.
Your particular item is not yet verified. Compare the identifiers and exclusions in the notice before deciding whether it applies.Agency-reported hazard: Ankle implantation systems were packaged with incorrect patient-specific cutting guides, which may result in: delayed surgery, the need to use standard instrumentation, reduced fixation or poor alignment of implants, implant loosening/migration, or need for revision surgery. Identifying details: System Serial Numbers: 18169, 18059. REF: 5831-9006, Lot: D-788, UDI-DI: 00840097509945.. Units affected: 2. Status: Ongoing.
1. Compare your item
This notice names a product; your particular item is not yet verified. Compare every applicable model, serial, lot, production-date or model-year boundary below.
Product named
Kinos Axiom Total Ankle System with Axiom PSR
Company / brand
restor3d
Lot / serial / date codes
System Serial Numbers: 18169, 18059. REF: 5831-9006, Lot: D-788, UDI-DI: 00840097509945.
Distribution
US Nationwide distribution in the states of FL, CO.
Agency product description and boundaries
Kinos Axiom Total Ankle System with Axiom PSR, Flat-cut (Cut Guide), REF: 5831-9006
Still to verify: match the label on your item with the full official notice, including any exclusions, tables and photos. Some identifiers appear only on the agency website and are absent from this data feed. A missing field does not mean all versions are affected.
If your lot or date falls outside a stated range, this notice has not established that your item is affected. This is not a finding about other recalls or product safety.
Ankle implantation systems were packaged with incorrect patient-specific cutting guides, which may result in: delayed surgery, the need to use standard instrumentation, reduced fix…