Atellica CH Enzymatic Hemoglobin A1c (A1c_E): recall status
Notice names this product. Siemens Healthcare Diagnostics’s Atellica CH Enzymatic Hemoglobin A1c (A1c_E) is named in FDA Class II recall Z-2703-2026.
Your particular item is not yet verified. Compare the identifiers and exclusions in the notice before deciding whether it applies.Agency-reported hazard: The potential for intermittent well-to-well bias on the A1c_E assay on the Atellica CH and Atellica CI analyzer which could cause falsely depressed or falsely elevated HbA1c patient results. Identifying details: Material Number: 11097536. UDI-DI: 00630414220505. All Lot Numbers.. Units affected: 24,077 units. Status: Ongoing.
1. Compare your item
This notice names a product; your particular item is not yet verified. Compare every applicable model, serial, lot, production-date or model-year boundary below.
Product named
Atellica CH Enzymatic Hemoglobin A1c (A1c_E)
Company / brand
Siemens Healthcare Diagnostics
Lot / serial / date codes
Material Number: 11097536. UDI-DI: 00630414220505. All Lot Numbers.
Distribution
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Bahamas, Bangladesh, Bolivia, Brazil, Bulgaria, Burkina¿Faso, Canada, Chile, China, Colombia, Croatia, Cura¿ao,¿St¿Eus, Czech¿Republic, Democratic Republic Congo, Denmark, Ecuador, Egypt, El Salvador, Estonia, Finland, France, Georgia, Germany, Greece, Hong Kong, India, Indonesia, Israel, Italy, Japan, Jordan, Kenya, Kuwait, Latvia, Lebanon, Lithuania, Macedonia, Malaysia, Mexico, Netherlands, New Zealand, Norway, Oman, Pakistan, Paraguay, Poland, Portugal, Qatar, Republic of Korea, Romania, Saudi¿Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Switzerland, Thailand, Turkey, Turkmenistan, U.A.E., Uganda, United¿Kingdom, Uruguay, Vatican City, Vietnam.
Agency product description and boundaries
Atellica CH Enzymatic Hemoglobin A1c (A1c_E). Material Number: 11097536.
Still to verify: match the label on your item with the full official notice, including any exclusions, tables and photos. Some identifiers appear only on the agency website and are absent from this data feed. A missing field does not mean all versions are affected.
If your lot or date falls outside a stated range, this notice has not established that your item is affected. This is not a finding about other recalls or product safety.
The potential for intermittent well-to-well bias on the A1c_E assay on the Atellica CH and Atellica CI analyzer which could cause falsely depressed or falsely elevated HbA1c patien…