Notice names this product. Siemens Healthcare Diagnostics’s Beckman Coulter is named in FDA Class II recall Z-2976-2026.
Your particular item is not yet verified. Compare the identifiers and exclusions in the notice before deciding whether it applies.Agency-reported hazard: Due to unacceptable precision performance exhibited in certain lots of the Emit 2000 Quinidine Assay, which may lead to erroneously elevated quinidine measurements, creating risk of sub-optimal patient treatment, including potential life-threatening cardiac arrhythmias or inadequate antimalarial therapy. Identifying details: Lots: 2597, 2627, 2709/UDI: 00842768023293. Units affected: 55 units. Status: Ongoing.
1. Compare your item
This notice names a product; your particular item is not yet verified. Compare every applicable model, serial, lot, production-date or model-year boundary below.
Product named
Beckman Coulter
Company / brand
Siemens Healthcare Diagnostics
Lot / serial / date codes
Lots: 2597, 2627, 2709/UDI: 00842768023293
Distribution
Worldwide - US Nationwide distribution in the state of CA and the country of Italy.
Agency product description and boundaries
Beckman Coulter, Emit 2000 Quinidine Assay REF: OSR4Q229 Contents: R1: 2 x 23 mL;R2 2x13 mL
Still to verify: match the label on your item with the full official notice, including any exclusions, tables and photos. Some identifiers appear only on the agency website and are absent from this data feed. A missing field does not mean all versions are affected.
If your lot or date falls outside a stated range, this notice has not established that your item is affected. This is not a finding about other recalls or product safety.
Due to unacceptable precision performance exhibited in certain lots of the Emit 2000 Quinidine Assay, which may lead to erroneously elevated quinidine measurements, creating risk o…