Fresh Recalls

Siemens Dimension Tacrolimus Flex Reagent Cartridge (TAC): recall status

Notice names this product. Siemens Healthcare Diagnostics’s Siemens Dimension Tacrolimus Flex Reagent Cartridge (TAC) is named in FDA Class II recall Z-2398-2026. Your particular item is not yet verified. Compare the identifiers and exclusions in the notice before deciding whether it applies. Agency-reported hazard: A potential positive bias in quality control (QC) and patient sample results in three lots of the Dimension Tacrolimus (TAC) Flex reagent cartridge. Identifying details: Manufacturing Lot/Serial Numbers: FA6254, FA6260, FA6286; UDIs: (01)00842768035425(10)FA6254(17)20260911; (01)00842768035425(10)FA6260(17)20260917; (01)00842768035425(10)FA6286(17)20261013.. Units affected: 2878 units. Status: Ongoing.

1. Compare your item

This notice names a product; your particular item is not yet verified. Compare every applicable model, serial, lot, production-date or model-year boundary below.

Product named
Siemens Dimension Tacrolimus Flex Reagent Cartridge (TAC)
Company / brand
Siemens Healthcare Diagnostics
Lot / serial / date codes
Manufacturing Lot/Serial Numbers: FA6254, FA6260, FA6286; UDIs: (01)00842768035425(10)FA6254(17)20260911; (01)00842768035425(10)FA6260(17)20260917; (01)00842768035425(10)FA6286(17)20261013.
Distribution
Worldwide distribution - US Nationwide and the countries of Argentina, Austria, Brazil, Croatia, Germany, Hungary, India, Italy, Japan, Korea, Lithuania, Pakistan, Poland, Saudi Arabia, Spain, South Africa, Sri Lanka, Turkey, United Arab Emirates.
Agency product description and boundaries

Siemens Dimension Tacrolimus Flex Reagent Cartridge (TAC), REF DF207, SMN 10700795; enzyme immunoassay

Still to verify: match the label on your item with the full official notice, including any exclusions, tables and photos. Some identifiers appear only on the agency website and are absent from this data feed. A missing field does not mean all versions are affected.

If your lot or date falls outside a stated range, this notice has not established that your item is affected. This is not a finding about other recalls or product safety.

Brand
Siemens Healthcare Diagnostics

Recalls naming this model

FDA Class II · Medical device · June 17, 2026

Siemens Dimension Tacrolimus Flex Reagent Cartridge (TAC) recall — Siemens Healthcare Diagnostics, Inc.

A potential positive bias in quality control (QC) and patient sample results in three lots of the Dimension Tacrolimus (TAC) Flex reagent cartridge.

Company / brand: Siemens Healthcare Diagnostics

Affected product: Siemens Dimension Tacrolimus Flex Reagent Cartridge (TAC)

Official ID Z-2398-2026 · Official record ↗

See all Siemens Healthcare Diagnostics recalls →

Verify with the official sources:

Found something wrong or missing? Report a data error.

Compiled from connected official public-domain government data. Sources last checked September 7, 2026 at 6:00 AM UTC.