Notice names this product. Siemens Medical Solutions USA’s ARTIS pheno is named in FDA Class II recall Z-2205-2026.
Your particular item is not yet verified. Compare the identifiers and exclusions in the notice before deciding whether it applies.Agency-reported hazard: During patient examinations, the system may sporadically display a 0 x-ray dose administered to the patient. This behavior has only been observed during fluoroscopy and/or single-frame acquisitions. Despite the reported x-ray dose showing 0, an actual x-ray dose is applied to the patient. Identifying details: UDI: 4056869046877. Status: Ongoing.
1. Compare your item
This notice names a product; your particular item is not yet verified. Compare every applicable model, serial, lot, production-date or model-year boundary below.
Product named
ARTIS pheno
Company / brand
Siemens Medical Solutions USA
Lot / serial / date codes
UDI: 4056869046877
Distribution
US and Worldwide
Agency product description and boundaries
ARTIS pheno, Model 10849000. interventional fluoroscopic x-ray system
Still to verify: match the label on your item with the full official notice, including any exclusions, tables and photos. Some identifiers appear only on the agency website and are absent from this data feed. A missing field does not mean all versions are affected.
If your lot or date falls outside a stated range, this notice has not established that your item is affected. This is not a finding about other recalls or product safety.
During patient examinations, the system may sporadically display a 0 x-ray dose administered to the patient. This behavior has only been observed during fluoroscopy and/or single-f…