TRIGEN PROXIMAL STRAIGHT HUMERAL NAIL 8/7 X 16: recall status
Notice names this product. Smith & Nephew’s TRIGEN PROXIMAL STRAIGHT HUMERAL NAIL 8/7 X 16 is named in FDA Class II recall Z-2456-2026.
Your particular item is not yet verified. Compare the identifiers and exclusions in the notice before deciding whether it applies.Agency-reported hazard: Humeral nail due to a manufacturing error is missing distal holes and if implanted may lead to revision surgery but, if the missing holes are identified before implantation, an equivalent backup can be used, or a bent nail of the same size, which would require additional surgical steps to address soft tissue/bone preparation, which could lead to a surgical delay of greater than 30 minutes. Identifying details: UDI-DI: 03596010514394. Lot/Batch: 25BM03616. Units affected: 5. Status: Ongoing.
1. Compare your item
This notice names a product; your particular item is not yet verified. Compare every applicable model, serial, lot, production-date or model-year boundary below.
Product named
TRIGEN PROXIMAL STRAIGHT HUMERAL NAIL 8/7 X 16
Company / brand
Smith & Nephew
Lot / serial / date codes
UDI-DI: 03596010514394. Lot/Batch: 25BM03616
Distribution
International distribution to the countries of BE, DE, IT.
Agency product description and boundaries
TRIGEN PROXIMAL STRAIGHT HUMERAL NAIL 8/7 X 16, REF: 71760816
Still to verify: match the label on your item with the full official notice, including any exclusions, tables and photos. Some identifiers appear only on the agency website and are absent from this data feed. A missing field does not mean all versions are affected.
If your lot or date falls outside a stated range, this notice has not established that your item is affected. This is not a finding about other recalls or product safety.
Humeral nail due to a manufacturing error is missing distal holes and if implanted may lead to revision surgery but, if the missing holes are identified before implantation, an equ…