Notice names this product. Staar Surgical AG’s Product Name is named in FDA Class II recall Z-2509-2026.
Your particular item is not yet verified. Compare the identifiers and exclusions in the notice before deciding whether it applies.Agency-reported hazard: Due to a software coding issues for calculating the implant orientation diagram printout for the implantation for toric implantable collamer lenses. Identifying details: Lot Code: UDI (01)00840311346288(0812)8.00 Software Version 8.00 The software error occurs if a customer generated an IOD for a toric implantation from April 13, 2026 to April 16, 2026 for multiple toric lens reservations for the same eye using the STAAR ICL Calculation Software.. Units affected: 543. Status: Ongoing.
1. Compare your item
This notice names a product; your particular item is not yet verified. Compare every applicable model, serial, lot, production-date or model-year boundary below.
Product named
Product Name
Company / brand
Staar Surgical AG
Lot / serial / date codes
Lot Code: UDI (01)00840311346288(0812)8.00 Software Version 8.00 The software error occurs if a customer generated an IOD for a toric implantation from April 13, 2026 to April 16, 2026 for multiple toric lens reservations for the same eye using the STAAR ICL Calculation Software.
Distribution
Worldwide - US Nationwide distribution including in the states and territory of AK, AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV and the countries of Austria, Australia, Belgium, Canada, Switzerland, Czech Republic, Germany Spain, Finland, France, Great Britain, Hong Kong, Ireland, Israel, India, Iran, Italy, Jordan, Korea, Kuwait, Kazakhstan, Morocco, Malaysia, Netherlands, Oman, Philippines, Poland, Saudi Arabia, Sweden, Singapore, Thailand, Turkey, Taiwan, Ukraine, USA, Vietnam.
Agency product description and boundaries
Product Name: STAAR ICL Calculation Software Software Version: Version 8.00 Product Description: The STAAR ICL Calculation Software is designed as an aide in the calculation of the size and diopter power (with residual refraction) for physician selection of the lens. For toric ICLs, an Implantation Orientation Diagram (IOD) is also produced to provide the physician with pictorial representation of the axis of rotation and alignment. Note: The STAAR ICL Calculation Software resides on STAAR Surgical's e-commerce website in STAAR ICL planning support (Stella), as referenced in the attached recall communication. Component: No
Still to verify: match the label on your item with the full official notice, including any exclusions, tables and photos. Some identifiers appear only on the agency website and are absent from this data feed. A missing field does not mean all versions are affected.
If your lot or date falls outside a stated range, this notice has not established that your item is affected. This is not a finding about other recalls or product safety.