Fresh Recalls

⚠ Data freshness notice: the most recent refresh of NHTSA was incomplete, so a very recent recall may be missing. The stored records may also have changed at the agency since the last successful check; for anything time-critical, confirm with the official sources linked below.

Techno-Path Multichem AE Unassayed Level 2 REF: recall status

Notice names this product. TECHNO-PATH MANUFACTURING’s Techno-Path Multichem AE Unassayed Level 2 REF is named in FDA Class II recall Z-0206-2027. Your particular item is not yet verified. Compare the identifiers and exclusions in the notice before deciding whether it applies. Agency-reported hazard: A routine stability studies have identified a stability drift affecting the Ammonia (NH4+) analyte in specific lots of Multichem AE quality control material. Additional testing confirmed that the labelled shelf-life performance claim for the Ammonia analyte can no longer be supported for the affected lots. This issue is limited to the Ammonia analyte only. The impacted device may cause delay of issuing results to the physician. Delayed recognition and diagnosis of elevated levels may impact in timely treatment of life-threatening levels described within medical toxicity intervention decision points. Identifying details: UDI:(01)05391523442817/Lot # 65705262. Units affected: 123 units. Status: Ongoing.

1. Compare your item

This notice names a product; your particular item is not yet verified. Compare every applicable model, serial, lot, production-date or model-year boundary below.

Product named
Techno-Path Multichem AE Unassayed Level 2 REF
Company / brand
TECHNO-PATH MANUFACTURING
Lot / serial / date codes
UDI:(01)05391523442817/Lot # 65705262
Distribution
US Nationwide distribution in the state of Massachusetts.
Agency product description and boundaries

Techno-Path Multichem AE Unassayed Level 2 REF: AE622X

Still to verify: match the label on your item with the full official notice, including any exclusions, tables and photos. Some identifiers appear only on the agency website and are absent from this data feed. A missing field does not mean all versions are affected.

If your lot or date falls outside a stated range, this notice has not established that your item is affected. This is not a finding about other recalls or product safety.

Brand
TECHNO-PATH MANUFACTURING

Recalls naming this model

FDA Class II · Medical device · October 28, 2026

Techno-Path Multichem AE Unassayed Level 2 REF recall — TECHNO-PATH MANUFACTURING LTD.

A routine stability studies have identified a stability drift affecting the Ammonia (NH4+) analyte in specific lots of Multichem AE quality control material. Additional testing con…

Company / brand: TECHNO-PATH MANUFACTURING recalls

Affected product: Techno-Path Multichem AE Unassayed Level 2 REF recall status

Official ID Z-0206-2027 · Official record ↗

See all TECHNO-PATH MANUFACTURING recalls →

Verify with the official sources:

Found something wrong or missing? Report a data error.

Compiled from connected official public-domain government data. Sources last checked November 25, 2026 at 12:00 AM UTC.