Notice names this product. Terumo Medical’s GLIDESHEATH SLENDER A-KIT 6F 10CM is named in FDA Class II recall Z-2930-2026.
Your particular item is not yet verified. Compare the identifiers and exclusions in the notice before deciding whether it applies.Agency-reported hazard: Product was packaged with the 0.025 guidewire instead of the 0.021 guidewire. Identifying details: Product Code: 60-1060. Shelfbox UDI-DI: 30389701011975; Pouch UDI-DI: 00389701011974; UDI-PI: (17)280501(10)0001441871. Lot Number: 0001441871.. Units affected: 7,490 units. Status: Ongoing.
1. Compare your item
This notice names a product; your particular item is not yet verified. Compare every applicable model, serial, lot, production-date or model-year boundary below.
Worldwide distribution - US Nationwide and the country of Canada.
Agency product description and boundaries
GLIDESHEATH SLENDER A-KIT 6F 10CM .021GW. 5 units per box. Product Code: 60-1060. The Glidesheath Slender is used to facilitate placing a catheter through the skin into the radial artery.
Still to verify: match the label on your item with the full official notice, including any exclusions, tables and photos. Some identifiers appear only on the agency website and are absent from this data feed. A missing field does not mean all versions are affected.
If your lot or date falls outside a stated range, this notice has not established that your item is affected. This is not a finding about other recalls or product safety.