ST AIA-Pack Testosterone: recall status

Notice names this product. Tosoh Bioscience’s ST AIA-Pack Testosterone is named in FDA Class II recall Z-2746-2026. Your particular item is not yet verified. Compare the identifiers and exclusions in the notice before deciding whether it applies. Agency-reported hazard: The Test cup lot F level 1 low-quality control results values for the device were lower than expected on the Automated Immunoassay Analyzer (AIA) which could cause falsely elevated or falsely low total testosterone results. Such erroneous test results may lead to additional testing/diagnostic interventions and in limited cases, may lead to a delay in diagnosis and/or initiation of unnecessary or incorrect clinical treatment measures. Identifying details: UDI-DI: 14560189210032; Lot Numbers: FX12404, FY12407, FY12406, FZ12408. Units affected: 310 units. Status: Ongoing.

1. Compare your item

This notice names a product; your particular item is not yet verified. Compare every applicable model, serial, lot, production-date or model-year boundary below.

Product named
ST AIA-Pack Testosterone
Company / brand
Tosoh Bioscience
Lot / serial / date codes
UDI-DI: 14560189210032; Lot Numbers: FX12404, FY12407, FY12406, FZ12408
Distribution
Global Distribution. US Nationwide.
Agency product description and boundaries

ST AIA-Pack Testosterone, IN VITRO DIAGNOSTIC USE ONLY for the quantitative measurement of testosterone in human serum on specific Tosoh AIA System analyzers, Model Number 0025204

Still to verify: match the label on your item with the full official notice, including any exclusions, tables and photos. Some identifiers appear only on the agency website and are absent from this data feed. A missing field does not mean all versions are affected.

If your lot or date falls outside a stated range, this notice has not established that your item is affected. This is not a finding about other recalls or product safety.

Brand
Tosoh Bioscience

Recalls naming this model

FDA Class II · Medical device · July 22, 2026

ST AIA-Pack Testosterone recall — Tosoh Bioscience Inc

The Test cup lot F level 1 low-quality control results values for the device were lower than expected on the Automated Immunoassay Analyzer (AIA) which could cause falsely elevated…

Company / brand: Tosoh Bioscience recalls

Affected product: ST AIA-Pack Testosterone recall status

Official ID Z-2746-2026 · Official record ↗

See all Tosoh Bioscience recalls →

Verify with the official sources:

Found something wrong or missing? Report a data error.

Compiled from connected official public-domain government data. Sources last checked October 22, 2026 at 6:00 PM UTC.