Fresh Recalls

FDAClass II · Medical device · Recall Z-0354-2027 · Recalled November 11, 2026

TruSystem 7500 Hybrid Plus (FC) recall — Baxter Healthcare Corporation

Agency report date: November 11, 2026.

1. Compare your item

This notice names a product; your particular item is not yet verified. Compare every applicable model, serial, lot, production-date or model-year boundary below.

Product named
TruSystem 7500 Hybrid Plus (FC)
Company / brand
Baxter Healthcare
Lot / serial / date codes
Serial numbers: 107971925; UDI-DI number: 00887761974067
Distribution
Worldwide - US Nationwide distribution.
Agency product description and boundaries

TruSystem 7500 Hybrid Plus (FC), Fixed-Installation Operating Table Column, Model/Catalog Number: 1854087

Still to verify: match the label on your item with the full official notice, including any exclusions, tables and photos. Some identifiers appear only on the agency website and are absent from this data feed. A missing field does not mean all versions are affected.

If your lot or date falls outside a stated range, this notice has not established that your item is affected. This is not a finding about other recalls or product safety.

2. Agency hazard and consumer action

FDA reports: There is an issue with the main lift mechanism. The main lift may unexpectedly drop from its set position to its lowest position during use. In these cases, the tabletop could not be raised again after the drop occurred. The issue has been associated with unexpected wear of the main lift assembly, also known as the telescopic spindle. An uncontrolled drop of the surgical table may result in an abrupt patient displacement during surgery. Serious or critical adverse health consequences may occur, including traumatic injury, airway compromise, and surgical procedure complications, among others. Baxter has received three complaints related to this issue, with no reported patient injury.

No consumer remedy instructions are supplied in this record. Check the official notice and listed company contact for the agency’s current instructions.

Class II: may cause temporary or medically reversible health consequences.

Affected product: TruSystem 7500 Hybrid Plus (FC) recall status

3. Open the official notice or contact

Official recall ID
Z-0354-2027
Brand
All Baxter Healthcare recalls
Classification
Class II
Units affected
1 unit
Recall status
Ongoing
Initiated
Voluntary: Firm initiated

Affected products

Companies & origin

Recalling firm location
Deerfield, IL, United States
Firm address
1 Baxter Pkwy, 60015-4625

Identifying details (verbatim from the notice)

Product codes / lots
Serial numbers: 107971925; UDI-DI number: 00887761974067
Distribution
Worldwide - US Nationwide distribution.
Firm notified the public via
Letter
FDA event ID
100117
Recall initiated
September 29, 2026
FDA classified
November 3, 2026
FDA report date
November 11, 2026

More Baxter Healthcare recalls

Other medical device recalls from around the same time

Verify with the official sources:

Found something wrong or missing? Report a data error.

Source: U.S. Food & Drug Administration. Agency excerpts come from the public data feed and may omit tables or later changes; the linked government source is authoritative. Source last checked November 21, 2026 at 12:00 AM UTC. Record retrieved November 21, 2026 at 12:00 AM UTC.