Fresh Recalls

FDAClass II · Medical device · Recall Z-0356-2027 · Recalled November 11, 2026

Unity Knee ASC Modular Handle included in the Corin UNITY ASC INSTRUMENT SYSTEM recall — CORIN MEDICAL, LTD.

Agency report date: November 11, 2026.

1. Compare your item

This notice names a product; your particular item is not yet verified. Compare every applicable model, serial, lot, production-date or model-year boundary below.

Product named
Unity Knee ASC Modular Handle included in the Corin UNITY ASC INSTRUMENT SYSTEM
Company / brand
CORIN MEDICAL
Lot / serial / date codes
Part Number: 1232040. UDI-DI/GTIN; 05056139239656. Lot Code: 097921.
Distribution
Worldwide - US Nationwide distribution in the states of CA, GA, IL, MI, NC, RI, TX, UT and the countries of Belgium, France, Germany, Hungary.
Agency product description and boundaries

Unity Knee ASC Modular Handle included in the Corin UNITY ASC INSTRUMENT SYSTEM. Part Number: 1232040.

Still to verify: match the label on your item with the full official notice, including any exclusions, tables and photos. Some identifiers appear only on the agency website and are absent from this data feed. A missing field does not mean all versions are affected.

If your lot or date falls outside a stated range, this notice has not established that your item is affected. This is not a finding about other recalls or product safety.

2. Agency hazard and consumer action

FDA reports: Potential malfunction of the Modular Handles where the locking mechanism may fail. Any mating instruments already engaged to the handle will not be able to be disengaged, and the handle will be unable to fitted securely to any unengaged mating instruments.

No consumer remedy instructions are supplied in this record. Check the official notice and listed company contact for the agency’s current instructions.

Class II: may cause temporary or medically reversible health consequences.

Affected product: Unity Knee ASC Modular Handle included in the Corin UNITY ASC INSTRUMENT SYSTEM recall status

3. Open the official notice or contact

Official recall ID
Z-0356-2027
Brand
All CORIN MEDICAL recalls
Classification
Class II
Units affected
22 units
Recall status
Ongoing
Initiated
Voluntary: Firm initiated

Affected products

Companies & origin

Recalling firm location
Cirencester, United Kingdom
Firm address
Unit 1-57-8, Corinium Centre, N/A

Identifying details (verbatim from the notice)

Product codes / lots
Part Number: 1232040. UDI-DI/GTIN; 05056139239656. Lot Code: 097921.
Distribution
Worldwide - US Nationwide distribution in the states of CA, GA, IL, MI, NC, RI, TX, UT and the countries of Belgium, France, Germany, Hungary.
Firm notified the public via
Letter
FDA event ID
99959
Recall initiated
October 2, 2026
FDA classified
November 3, 2026
FDA report date
November 11, 2026