FDAClass II · Medical device · Recalled May 13, 2026
Brand Name recall — Intuitive Surgical, Inc.
Hazard: Due to manipulator containing screws with in the arm sub-assembly that my be susceptible to breaking.
Class II: may cause temporary or medically reversible health consequences.
View the official FDA enforcement record (openFDA) ↗
Description
Brand Name: Da Vinci X and Da Vinci Xi Product Name: da Vinci X, Xi Surgical System Model/Catalog Number: IS4000; IS4200 Software Version: N/A Product Description: he da Vinci Xi, X Surgical systems each consist of three main components: the Surgeon Console (SC), the Vision Cart (VC), and the Patient Cart (PC). The PC is the operative component of the system, and its primary function is to support the instrument arms and camera arm. Each system have four Universal Surgical Manipulators (USMs) also known as arms, which allow the surgeon to control up to three instruments and an endoscope simultaneously to perform surgical tasks (e.g. cutting, suturing, tissue approximation) in a variety of procedures. As the surgeon moves the hand controls within the console, the PC moves the robotic instruments within the patient in a corresponding manner. The da Vinci system and its software are designed such that instruments motion precisely tracks the motion inputs from the hand controls during surgery. Component: The affected part based on this recall, 380647-xx, USM, is a sub component of da Vinci Surgical Instrument
Affected products
Companies & origin
Identifying details (verbatim from the notice)
Source: U.S. Food & Drug Administration. This page mirrors the official public-domain record verbatim; the linked government source is authoritative. Source last checked July 24, 2026 at 3:00 AM UTC.