Intuitive Surgical recalls
5 recent recalls on record for Intuitive Surgical in connected official government data as of October 22, 2026 at 6:00 PM UTC.
Models with recalls
All Intuitive Surgical recalls
FDA Class II · Medical device · July 22, 2026
Robotic-assisted surgical system arm drapes may detach from the instrument arm due to an issue involving the drape rings located at the top of the product, which may expose a secti…
Official ID Z-2722-2026 ·
Official record ↗
FDA Class II · Medical device · July 1, 2026
Operative components of surgical systems potentially distributed with E-brake retainer pin, so in event of a structural column failure, the E-brake pin would prevent E-brake engage…
Official ID Z-2521-2026 ·
Official record ↗
FDA Class II · Medical device · June 10, 2026
Due to unintentionally activated near infrared handheld camera when attempting to turn off illumination during normal clinical workflow transitions, particularly when switching fro…
Official ID Z-2310-2026 ·
Official record ↗
FDA Class II · Medical device · May 13, 2026
Due to manipulator containing screws with in the arm sub-assembly that my be susceptible to breaking.
Official ID Z-2046-2026 ·
Official record ↗
FDA Class I · Medical device · May 13, 2026
Reloads for curved-tipped stapler may potentially produce an incomplete staple line.
Official ID Z-1855-2026 ·
Official record ↗