Fresh Recalls

FDAClass II · Medical device · Recall Z-2722-2026 · Recalled July 22, 2026

da Vinci SP Instrument Arm Drape recall — Intuitive Surgical, Inc.

Agency report date: July 22, 2026.

1. Compare your item

This notice names a product; your particular item is not yet verified. Compare every applicable model, serial, lot, production-date or model-year boundary below.

Product named
da Vinci SP Instrument Arm Drape
Company / brand
Intuitive Surgical
Lot / serial / date codes
UDI-DI: 00886874113486, Lot: 3074LA0500, 1124LA0200, 2124LA0400, 1134LA0700, 1144LA0500, 7154LA0200, 6164LA0800, 1084LA0300, 3164LAA800, 4084LA0500, 4124LA0600, 1274LA0500, 2294LA0200, 1274LA0400, 4284LA0800, 6324LA0200, 1324LA0300, 4154LA0400, 2284LA0100, 4144LA1300, 3324LA0500, 2354LA0500, 7144LA0500, 4294LA0700, 3374LA0400, 5264LA0700, 3224LA0500, 3424LA0200, 1394LA0100
Distribution
Worldwide - US Nationwide distribution in the states of NJ, OR, NY, FL, AL, SD, TX, VA, WV, PA, CA, WA, CT, MA, IL, OK, OH, MI, NC, MO, KY, WI, AZ, CO, GA, MD, SC, IN, NV, TN, MS, KS, IA, UT, MN, LA, NE, ME and the countries of Germany, Sweden, United Kingdom, Italy, Belgium, Switzerland, Denmark, Japan, South Korea, Hong Kong, Taiwan, France, Ireland, Spain.
Agency product description and boundaries

da Vinci SP Instrument Arm Drape, SP, 1-PACK, Part Number: 430013-13, 430013-15

Still to verify: match the label on your item with the full official notice, including any exclusions, tables and photos. Some identifiers appear only on the agency website and are absent from this data feed. A missing field does not mean all versions are affected.

If your lot or date falls outside a stated range, this notice has not established that your item is affected. This is not a finding about other recalls or product safety.

2. Agency hazard and consumer action

FDA reports: Robotic-assisted surgical system arm drapes may detach from the instrument arm due to an issue involving the drape rings located at the top of the product, which may expose a section of instrument arm, which may result in procedure delay or conversion to an alternate minimally invasive approach. Retrospectively reported recall.

No consumer remedy instructions are supplied in this record. Check the official notice and listed company contact for the agency’s current instructions.

Class II: may cause temporary or medically reversible health consequences.

Affected product: da Vinci SP Instrument Arm Drape recall status

3. Open the official notice or contact

Official recall ID
Z-2722-2026
Brand
All Intuitive Surgical recalls
Classification
Class II
Units affected
11,113
Recall status
Ongoing
Initiated
Voluntary: Firm initiated

Affected products

Companies & origin

Recalling firm location
Sunnyvale, CA, United States
Firm address
1266 Kifer Rd, 94086-5304

Identifying details (verbatim from the notice)

Product codes / lots
UDI-DI: 00886874113486, Lot: 3074LA0500, 1124LA0200, 2124LA0400, 1134LA0700, 1144LA0500, 7154LA0200, 6164LA0800, 1084LA0300, 3164LAA800, 4084LA0500, 4124LA0600, 1274LA0500, 2294LA0200, 1274LA0400, 4284LA0800, 6324LA0200, 1324LA0300, 4154LA0400, 2284LA0100, 4144LA1300, 3324LA0500, 2354LA0500, 7144LA0500, 4294LA0700, 3374LA0400, 5264LA0700, 3224LA0500, 3424LA0200, 1394LA0100
Distribution
Worldwide - US Nationwide distribution in the states of NJ, OR, NY, FL, AL, SD, TX, VA, WV, PA, CA, WA, CT, MA, IL, OK, OH, MI, NC, MO, KY, WI, AZ, CO, GA, MD, SC, IN, NV, TN, MS, KS, IA, UT, MN, LA, NE, ME and the countries of Germany, Sweden, United Kingdom, Italy, Belgium, Switzerland, Denmark, Japan, South Korea, Hong Kong, Taiwan, France, Ireland, Spain.
FDA event ID
98890
Recall initiated
May 12, 2025
FDA classified
July 10, 2026
FDA report date
July 22, 2026

More Intuitive Surgical recalls

Other medical device recalls from around the same time

Verify with the official sources:

Found something wrong or missing? Report a data error.

Source: U.S. Food & Drug Administration. Agency excerpts come from the public data feed and may omit tables or later changes; the linked government source is authoritative. Source last checked October 22, 2026 at 6:00 PM UTC. Record retrieved October 20, 2026 at 6:00 PM UTC.

Report a data error on this page