Fresh Recalls

FDAClass II · Medical device · Recall Z-3088-2026 · Recalled September 9, 2026

Velys Enabling Tech Robotic-Assist Station REF recall — Medos International Sarl

Agency report date: September 9, 2026.

1. Compare your item

This notice names a product; your particular item is not yet verified. Compare every applicable model, serial, lot, production-date or model-year boundary below.

Product named
Velys Enabling Tech Robotic-Assist Station REF
Company / brand
Medos International Sarl
Lot / serial / date codes
UDI: 10705034645995/Serial Numbers: W14316295301 W15316297501 W15316298001 W17316322901 W17316323301 W21316347401 W21316347501 W21316347601 W21316347801 W21316347901 W21316348001 W22316365801 W34316479001 W34316479101 W34316479201 W34316479301 W34316479401 W34316479601 W34316479801 W34316480001 W34316480101 W49316554301 W51316597001 W51316597101 Y03316603601 Y03316603701 Y03316603801 Y07316625901 Y15316690901
Distribution
US Nationwide distribution in the states of DE, FL, IL, KY, MD, MI, MN, NJ, NM, NY, OR, PA, TN, TX, VA, WA.
Agency product description and boundaries

Velys Enabling Tech Robotic-Assist Station REF: 451611001

Still to verify: match the label on your item with the full official notice, including any exclusions, tables and photos. Some identifiers appear only on the agency website and are absent from this data feed. A missing field does not mean all versions are affected.

If your lot or date falls outside a stated range, this notice has not established that your item is affected. This is not a finding about other recalls or product safety.

2. Agency hazard and consumer action

FDA reports: Software issue within robotic controller which may cause unintended movement of the robotic arm during bone preparation or instrument positioning.

No consumer remedy instructions are supplied in this record. Check the official notice and listed company contact for the agency’s current instructions.

Class II: may cause temporary or medically reversible health consequences.

Affected product: Velys Enabling Tech Robotic-Assist Station REF recall status

3. Open the official notice or contact

Official recall ID
Z-3088-2026
Brand
All Medos International Sarl recalls
Classification
Class II
Units affected
29 stations
Recall status
Ongoing
Initiated
Voluntary: Firm initiated

Affected products

Companies & origin

Recalling firm location
Le Locle, N/A, Switzerland
Firm address
Chemin Blanc 36, N/A, N/A

Identifying details (verbatim from the notice)

Product codes / lots
UDI: 10705034645995/Serial Numbers: W14316295301 W15316297501 W15316298001 W17316322901 W17316323301 W21316347401 W21316347501 W21316347601 W21316347801 W21316347901 W21316348001 W22316365801 W34316479001 W34316479101 W34316479201 W34316479301 W34316479401 W34316479601 W34316479801 W34316480001 W34316480101 W49316554301 W51316597001 W51316597101 Y03316603601 Y03316603701 Y03316603801 Y07316625901 Y15316690901
Distribution
US Nationwide distribution in the states of DE, FL, IL, KY, MD, MI, MN, NJ, NM, NY, OR, PA, TN, TX, VA, WA.
Firm notified the public via
Visit
FDA event ID
99467
Recall initiated
July 23, 2026
FDA classified
September 2, 2026
FDA report date
September 9, 2026

Other medical device recalls from around the same time

Verify with the official sources:

Found something wrong or missing? Report a data error.

Source: U.S. Food & Drug Administration. Agency excerpts come from the public data feed and may omit tables or later changes; the linked government source is authoritative. Source last checked October 22, 2026 at 6:00 PM UTC. Record retrieved October 22, 2026 at 6:00 PM UTC.

Report a data error on this page