FDAClass II · Medical device · Recall Z-3191-2026 · Recalled September 23, 2026

Aisys CS2 with Et Control recall — Datex-Ohmeda, Inc.

Agency report date: September 23, 2026.

1. Compare your item

This notice names a product; your particular item is not yet verified. Compare every applicable model, serial, lot, production-date or model-year boundary below.

Product named
Aisys CS2 with Et Control
Company / brand
Datex-Ohmeda
Lot / serial / date codes
Serial Numbers: ANAS01357 ANAP00609 ANAQ00988 ANAS00586 ANAR00543 ANAP01088 ANAM00443
Distribution
Worldwide - US Nationwide distribution.
Agency product description and boundaries

Aisys CS2 with Et Control, Datex-Ohmeda anesthesia system, Part Number 1011-9000-000

Still to verify: match the label on your item with the full official notice, including any exclusions, tables and photos. Some identifiers appear only on the agency website and are absent from this data feed. A missing field does not mean all versions are affected.

If your lot or date falls outside a stated range, this notice has not established that your item is affected. This is not a finding about other recalls or product safety.

2. Agency hazard and consumer action

FDA reports: Anesthesia Delivery Systems containing certain electronic gas mixer assemblies can automatically transition to Alternate O2 control. In Alternate O2 control, the system delivers 100% O2 to the breathing system enabling continued controlled or manual ventilation and initiates audible and visual alarms. When this issue occurs, anesthetic agent delivery will be interrupted and requires clinician intervention. If this is not provided, the patient may experience decreased anesthetic agent effect.

No consumer remedy instructions are supplied in this record. Check the official notice and listed company contact for the agency’s current instructions.

Class II: may cause temporary or medically reversible health consequences.

Affected product: Aisys CS2 with Et Control recall status

3. Open the official notice or contact

Official recall ID
Z-3191-2026
Brand
All Datex-Ohmeda recalls
Classification
Class II
Units affected
7 units
Recall status
Ongoing
Initiated
Voluntary: Firm initiated

Affected products

Companies & origin

Recalling firm location
Madison, WI, United States
Firm address
3030 Ohmeda Dr, N/A, 53718-6704

Identifying details (verbatim from the notice)

Product codes / lots
Serial Numbers: ANAS01357 ANAP00609 ANAQ00988 ANAS00586 ANAR00543 ANAP01088 ANAM00443
Distribution
Worldwide - US Nationwide distribution.
Firm notified the public via
Letter
FDA event ID
99649
Recall initiated
July 30, 2026
FDA classified
September 16, 2026
FDA report date
September 23, 2026