Fresh Recalls

FDAClass II · Medical device · Recall Z-3193-2026 · Recalled September 23, 2026

Captus 4000e Thyroid Uptake System recall — Mirion Technologies (Capintec), Inc.

Agency report date: September 23, 2026.

1. Compare your item

This notice names a product; your particular item is not yet verified. Compare every applicable model, serial, lot, production-date or model-year boundary below.

Product named
Captus 4000e Thyroid Uptake System
Company / brand
Mirion Technologies (Capintec)
Lot / serial / date codes
Affected Units: Captus 4000: 1. Model 5430-30151; UDI-DI: 00859942006096; 2. Model 5430-30152; UDI-DI: 00859942006102; 3. Model 5430-30154; UDI-DI: 00859942006423; Devices shipped from April 2019 through February 2024, Serial Numbers 940548 through 941398, that have NOT already replaced the original arm with the redesigned arm.
Distribution
Worldwide distribution - US Nationwide and the countries of ALGERIA, AUSTRIA, BANGLADESH, BRASIL, Brazil, CANADA, CHILE, Costa Rica, Ecuador, France, Hanoi Vietnam, Hong Kong, INDIA, ISRAEL, ITALY, Malaysia, Philippines, POLAND, Saudi Arabia, SOUTH KOREA, Spain, SWEDEN, TAIWAN, THAILAND, TURKEY, UNITED ARAB EMIATES, UNITED KINGDOM, VIETNAM.
Agency product description and boundaries

Captus 4000e Thyroid Uptake System, Models: 1. 5430-30151, 2. 5430-30152, 3. 5430-30154;

Still to verify: match the label on your item with the full official notice, including any exclusions, tables and photos. Some identifiers appear only on the agency website and are absent from this data feed. A missing field does not mean all versions are affected.

If your lot or date falls outside a stated range, this notice has not established that your item is affected. This is not a finding about other recalls or product safety.

2. Agency hazard and consumer action

FDA reports: Potential for failure of the tension bolt which would allow the spring arm to unexpectedly fall to its lowest position, which is approximately 25 inches (63.5 cm) from the ground.

No consumer remedy instructions are supplied in this record. Check the official notice and listed company contact for the agency’s current instructions.

Class II: may cause temporary or medically reversible health consequences.

Affected product: Captus 4000e Thyroid Uptake System recall status

3. Open the official notice or contact

Official recall ID
Z-3193-2026
Brand
All Mirion Technologies (Capintec) recalls
Classification
Class II
Units affected
783
Recall status
Ongoing
Initiated
Voluntary: Firm initiated

Affected products

Companies & origin

Recalling firm location
Florham Park, NJ, United States
Firm address
7 Vreeland Rd, N/A, 07932-1511

Identifying details (verbatim from the notice)

Product codes / lots
Affected Units: Captus 4000: 1. Model 5430-30151; UDI-DI: 00859942006096; 2. Model 5430-30152; UDI-DI: 00859942006102; 3. Model 5430-30154; UDI-DI: 00859942006423; Devices shipped from April 2019 through February 2024, Serial Numbers 940548 through 941398, that have NOT already replaced the original arm with the redesigned arm.
Distribution
Worldwide distribution - US Nationwide and the countries of ALGERIA, AUSTRIA, BANGLADESH, BRASIL, Brazil, CANADA, CHILE, Costa Rica, Ecuador, France, Hanoi Vietnam, Hong Kong, INDIA, ISRAEL, ITALY, Malaysia, Philippines, POLAND, Saudi Arabia, SOUTH KOREA, Spain, SWEDEN, TAIWAN, THAILAND, TURKEY, UNITED ARAB EMIATES, UNITED KINGDOM, VIETNAM.
Firm notified the public via
E-Mail
FDA event ID
99677
Recall initiated
July 20, 2026
FDA classified
September 16, 2026
FDA report date
September 23, 2026

Other medical device recalls from around the same time

Verify with the official sources:

Found something wrong or missing? Report a data error.

Source: U.S. Food & Drug Administration. Agency excerpts come from the public data feed and may omit tables or later changes; the linked government source is authoritative. Source last checked October 22, 2026 at 6:00 PM UTC. Record retrieved October 22, 2026 at 6:00 PM UTC.

Report a data error on this page