Fresh Recalls

FDAClass II · Medical device · Recall Z-3257-2026 · Recalled September 30, 2026

Brand Name recall — Baxter Healthcare Corporation

Agency report date: September 30, 2026.

1. Compare your item

This notice names a product; your particular item is not yet verified. Compare every applicable model, serial, lot, production-date or model-year boundary below.

Product named
Brand Name
Company / brand
Baxter Healthcare
Lot / serial / date codes
Lot Code: UDI: 00887761981690 Serial Numbers: W078SY3553 and Z309SY7853
Distribution
US Nationwide distribution in the states of Colorado and Florida.
Agency product description and boundaries

Brand Name: VOLARA System Product Name: VOLARA, HC, NA A Model/Catalog Number: M08594A Software Version: N/A Product Description: VOLARA, HOMECARE, NORTH AMERICA Component: No

Still to verify: match the label on your item with the full official notice, including any exclusions, tables and photos. Some identifiers appear only on the agency website and are absent from this data feed. A missing field does not mean all versions are affected.

If your lot or date falls outside a stated range, this notice has not established that your item is affected. This is not a finding about other recalls or product safety.

2. Agency hazard and consumer action

FDA reports: Contamination within the release test equipment flow path could have affected the accuracy of the nebulizer flow rate. Affected devices returned to firm and testing confirmed devices meet specifications.

No consumer remedy instructions are supplied in this record. Check the official notice and listed company contact for the agency’s current instructions.

Class II: may cause temporary or medically reversible health consequences.

Affected product: Brand Name recall status

3. Open the official notice or contact

Official recall ID
Z-3257-2026
Brand
All Baxter Healthcare recalls
Classification
Class II
Units affected
2
Recall status
Ongoing
Initiated
Voluntary: Firm initiated

Affected products

Companies & origin

Recalling firm location
Deerfield, IL, United States
Firm address
1 Baxter Pkwy, 60015-4625

Identifying details (verbatim from the notice)

Product codes / lots
Lot Code: UDI: 00887761981690 Serial Numbers: W078SY3553 and Z309SY7853
Distribution
US Nationwide distribution in the states of Colorado and Florida.
Firm notified the public via
Telephone
FDA event ID
99681
Recall initiated
August 3, 2026
FDA classified
September 23, 2026
FDA report date
September 30, 2026

More Baxter Healthcare recalls

Other medical device recalls from around the same time

Verify with the official sources:

Found something wrong or missing? Report a data error.

Source: U.S. Food & Drug Administration. Agency excerpts come from the public data feed and may omit tables or later changes; the linked government source is authoritative. Source last checked October 22, 2026 at 12:00 PM UTC. Record retrieved October 22, 2026 at 12:00 PM UTC.