Fresh Recalls

Brand Name: recall status

Notice names this product. Baxter Healthcare’s Brand Name is named in FDA Class II recall Z-2262-2026. Your particular item is not yet verified. Compare the identifiers and exclusions in the notice before deciding whether it applies. Agency-reported hazard: Affected lots may contain incorrect 0.8-micron filters. Use of a 0.8-micron filter in place of the validated 3-micron filter may lead to reduced flow efficiency, incomplete or slower transfers of ingredients, or system alarms. This issue may lead to clinical effects such as electrolyte imbalance, metabolic instability, delayed recovery, etc. Identifying details: Lot Code: Lot numbers: 804084, 804088, and 804089 UDI Number: 00085412475806. Units affected: 105,675. Status: Ongoing.

1. Compare your item

This notice names a product; your particular item is not yet verified. Compare every applicable model, serial, lot, production-date or model-year boundary below.

Product named
Brand Name
Company / brand
Baxter Healthcare
Lot / serial / date codes
Lot Code: Lot numbers: 804084, 804088, and 804089 UDI Number: 00085412475806
Distribution
Worldwide - US Nationwide distribution including in the states of Alabama, Arkansas, Arizona, California, Colorado, Florida, Georgia, Hawaii, Iowa, Idaho, Illinois, Indiana, Kansas, Kentucky, Louisiana, Massachusetts, Maine, Maryland, Michigan, Minnesota, Missouri, Mississippi, Montana, North Carolina, North Dakota, Nebraska, New Hampshire, New Jersey, New Mexico, Nevada, New York, Ohio, Oklahoma, Oregon, Pennsylvania, South Dakota, Tennessee, Texas, Utah, Virginia, Washington, Wisconsin, and West Virginia. The countries of Taiwan and Uruguay.
Agency product description and boundaries

Brand Name: Exacta Mix Product Name: Vented Micro-Volume Inlet Model/Catalog Number: H938175 Product Description: Tubing sets with push-on connector on one end (to attach to the Fluid Selector Valve) and a spike or other appropriate connector on the opposite end to attach the source container.

Still to verify: match the label on your item with the full official notice, including any exclusions, tables and photos. Some identifiers appear only on the agency website and are absent from this data feed. A missing field does not mean all versions are affected.

If your lot or date falls outside a stated range, this notice has not established that your item is affected. This is not a finding about other recalls or product safety.

Brand
Baxter Healthcare

Recalls naming this model

FDA Class II · Medical device · June 10, 2026

Brand Name recall — Baxter Healthcare Corporation

Affected lots may contain incorrect 0.8-micron filters. Use of a 0.8-micron filter in place of the validated 3-micron filter may lead to reduced flow efficiency, incomplete or slow…

Company / brand: Baxter Healthcare

Affected product: Brand Name

Official ID Z-2262-2026 · Official record ↗

See all Baxter Healthcare recalls →