Fresh Recalls

Beckman Coulter DxC 500 AU Clinical Analyzer Modules using software version 1: recall status

Notice names this product. Beckman Coulter’s Beckman Coulter DxC 500 AU Clinical Analyzer Modules using software version 1 is named in FDA Class II recall Z-2974-2026. Your particular item is not yet verified. Compare the identifiers and exclusions in the notice before deciding whether it applies. Agency-reported hazard: a software defect in which the Parameter Package (PP) import logic overwrites all predefined calculated test rules, including those previously modified by the user, upon import of a new Parameter Package version. As a result, usercustomized calculated test rules revert silently to their default expressions, potentially generating erroneous patient test results without an explicit error notification. The issue affects instruments running software version 1.5 or lower and only occurs when a new Parameter Package is imported to a system that already has an existing Parameter Package onboard; new installations are not affected. Identifying details: 1) Part Number C63519, UDI/DI 14987666545058 2) Part Number C63520, UDI/DI 14987666545065. Status: Ongoing.

1. Compare your item

This notice names a product; your particular item is not yet verified. Compare every applicable model, serial, lot, production-date or model-year boundary below.

Product named
Beckman Coulter DxC 500 AU Clinical Analyzer Modules using software version 1
Company / brand
Beckman Coulter
Lot / serial / date codes
1) Part Number C63519, UDI/DI 14987666545058 2) Part Number C63520, UDI/DI 14987666545065
Distribution
Worldwide distribution - US Nationwide.
Agency product description and boundaries

Beckman Coulter DxC 500 AU Clinical Analyzer Modules using software version 1.5 or lower: 1) AU w/out ISE, Part Number C63519; 2) AU w/ISE, Part Number C63520; DxC 500 AU Clinical Chemistry Analyzers

Still to verify: match the label on your item with the full official notice, including any exclusions, tables and photos. Some identifiers appear only on the agency website and are absent from this data feed. A missing field does not mean all versions are affected.

If your lot or date falls outside a stated range, this notice has not established that your item is affected. This is not a finding about other recalls or product safety.

Brand
Beckman Coulter

Recalls naming this model

FDA Class II · Medical device · August 26, 2026

Beckman Coulter DxC 500 AU Clinical Analyzer Modules using software version 1 recall — Beckman Coulter Inc.

a software defect in which the Parameter Package (PP) import logic overwrites all predefined calculated test rules, including those previously modified by the user, upon import of …

Company / brand: Beckman Coulter recalls

Affected product: Beckman Coulter DxC 500 AU Clinical Analyzer Modules using software version 1 recall status

Official ID Z-2974-2026 · Official record ↗

See all Beckman Coulter recalls →

Verify with the official sources:

Found something wrong or missing? Report a data error.

Compiled from connected official public-domain government data. Sources last checked October 22, 2026 at 6:00 PM UTC.