FDAClass II · Medical device · Recall Z-2974-2026 · Recalled August 26, 2026

Beckman Coulter DxC 500 AU Clinical Analyzer Modules using software version 1 recall — Beckman Coulter Inc.

Agency report date: August 26, 2026.

1. Compare your item

This notice names a product; your particular item is not yet verified. Compare every applicable model, serial, lot, production-date or model-year boundary below.

Product named
Beckman Coulter DxC 500 AU Clinical Analyzer Modules using software version 1
Company / brand
Beckman Coulter
Lot / serial / date codes
1) Part Number C63519, UDI/DI 14987666545058 2) Part Number C63520, UDI/DI 14987666545065
Distribution
Worldwide distribution - US Nationwide.
Agency product description and boundaries

Beckman Coulter DxC 500 AU Clinical Analyzer Modules using software version 1.5 or lower: 1) AU w/out ISE, Part Number C63519; 2) AU w/ISE, Part Number C63520; DxC 500 AU Clinical Chemistry Analyzers

Still to verify: match the label on your item with the full official notice, including any exclusions, tables and photos. Some identifiers appear only on the agency website and are absent from this data feed. A missing field does not mean all versions are affected.

If your lot or date falls outside a stated range, this notice has not established that your item is affected. This is not a finding about other recalls or product safety.

2. Agency hazard and consumer action

FDA reports: a software defect in which the Parameter Package (PP) import logic overwrites all predefined calculated test rules, including those previously modified by the user, upon import of a new Parameter Package version. As a result, usercustomized calculated test rules revert silently to their default expressions, potentially generating erroneous patient test results without an explicit error notification. The issue affects instruments running software version 1.5 or lower and only occurs when a new Parameter Package is imported to a system that already has an existing Parameter Package onboard; new installations are not affected.

No consumer remedy instructions are supplied in this record. Check the official notice and listed company contact for the agency’s current instructions.

Class II: may cause temporary or medically reversible health consequences.

Affected product: Beckman Coulter DxC 500 AU Clinical Analyzer Modules using software version 1 recall status

3. Open the official notice or contact

Official recall ID
Z-2974-2026
Brand
All Beckman Coulter recalls
Classification
Class II
Recall status
Ongoing
Initiated
Voluntary: Firm initiated

Affected products

Companies & origin

Recalling firm location
Brea, CA, United States
Firm address
250 S Kraemer Blvd, 92821-6232

Identifying details (verbatim from the notice)

Product codes / lots
1) Part Number C63519, UDI/DI 14987666545058 2) Part Number C63520, UDI/DI 14987666545065
Distribution
Worldwide distribution - US Nationwide.
Firm notified the public via
Letter
FDA event ID
99436
Recall initiated
July 15, 2026
FDA classified
August 17, 2026
FDA report date
August 26, 2026

Other medical device recalls from around the same time